Owner manual

Section 12: Specifications
05/11 96 51209 Revision 2
12.2 General Data
The following characteristics are valid after a warm-up time of at least 60 minutes (reference
temperature: 2C).
12.2.1 Power Supply
Supply voltage 24 VDC +/- 20 %
Power
consumption
6 Watts
Power
consumption
with fieldbus option 8 Watts
12.3 Effect of Ambient Conditions
12.3.1 Environmental Conditions
Temperature range
Reference temperature 23 °C
Ambient temperature for
operation
10,000 e -10... +40 °C
‘not legal for
trade’
-10... +50 °C
Power-on temperature 0... +50 °C
Storage/transport -40... +70 °C
Humidity < 95 %, no condensation, (acc. to IEC 68-2)
Protection type to DIN 40050 IP30
12.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial area
Housing Radio frequency electromagnetic fields (80 – 1000 MHz)
EN 61000-4-3 10 V/m
Radio frequency electromagnetic fields (1.4 – 2.0 GHz) EN 61000-4-3 3 V/m
Radio frequency electromagnetic fields (2.0 – 2.7 GHz) EN 61000-4-3 1 V/m
Electrostatic discharge (ESD) EN 61000-4-2 6 / 8 kV
Signal and Fast transients (burst) EN 61000-4-4 1 kV
control lines Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1 kV
Conducted disturbances by radio frequency (0.15 –
80 MHz)
EN 61000-4-6 10 V
Mains input Fast transients (burst) EN 61000-4-4 2 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1 / 2 kV
Conducted disturbances by radio frequency (0.15 –
80 MHz)
EN 61000-4-6 10 V
12.3.3 RF Interference Suppression
Electromagnetic emission In acc. with EN 61326, limit value class A, industrial area