User's Manual

WEMS Installer’s Guide 38
WEMS Software
iWPM-T-Specific Menus
The iWPM-T tabs include menus for calibrating the pressure and temperature
sensors on the iWPM-T device. Figure 27 below shows the menu for calibrating
the pressure sensor, and Figure 28 on the following page shows the temperature
sensor calibration menu. Luna iMonitoring calibrates all of the iWPM-T’s sensors
before shipping.
iWPM-T Pressure Sensor Calibration
To set the pressure calibration parameters, select the Pres Cal tab at the bottom
of the screen to access the pressure calibration menu shown in Figure 27 below.
Select the proper Units from the pull-down list. The choices for Units are
millivolts, PSI, mm Hg, kPa, and mBar. Enter the correct Slope and Intercept for
the specified units, and hit the Set Pressure Cal button.
Figure 27.
i
WPM-T Pressure Sensor Calibration Menu.