Users Manual Part 1

Getting started
30
Site requirements
Caution
z
Ensure that the laboratory floor is level and with a gradient of less than
1/200.
z
Ensure that the laboratory is free of dust, corrosive and flammable gas, and
heat and wind sources.
z
Ensure that the laboratory is away from direct sunlight and is well
ventilated. We recommend that you refer to the standard of a biosafety
level (BSL) 2 laboratory.
z
We recommend that you use a clean laboratory with ISO Class 8 air
cleanliness. Ensure that air can be circulated inside the laboratory but
cannot be circulated to other laboratories.
z
The device should be used with the ultra-pure water machine, refrigerator,
and DNB loader, and can be used with the library preparation system.
Therefore, enough space should be considered when placing the above
instruments.
z
Ensure that enough space is provided around the device for ventilation,
cable connection, and power switch operation.
The following figure indicates distances that are required for optimal
operation and access.
1500 mm (59.1 inches)
B
A
B
1000 mm (39.4 inches)
Operation area
B
B
A
Figure 13 Space requirements
Network requirements
Operating environments of the computer
Minimum configuration
Processor: Intel Core i7-7700