Operator's Manual

2524
Settings
Access SETTINGS from the top right corner of
the Job Menu. Each setting button cycles through
its options or to its own screen (blue arrow).
The DEFAULTS button in the top right corner of
the settings reverts to default settings, but jobs
are not deleted.
Job Link® System: OFF (default), ON
Select ON to view live combustion and saved tests on the Job Link
system mobile app. (Aects battery life.)
Brightness: HIGH, MEDIUM (default), LOW
(Aects battery life.)
Button Sound: ON (default), OFF
(Aects battery life.)
Sensor Remaining (status): % O2 and % CO sensor life remaining.
New sensors show 100%. When life of either sensor is below 5%, a
notice appears at startup.
Memory % Free (status): % of memory remaining for saving jobs.
When memory is too low, a notice appears when trying to SAVE AND
EXIT a job. DELETE ALL is a quick way to delete all jobs and restart at
JOB #1. Delete single jobs from within the JOB ARCHIVE screen.
Date and Time: EDIT to adjust date and time. 24 hour time is always
used. Date format is always set to Year-Month-Day for easy sorting.
Company Information: EDIT to adjust Company Information shown
on reports.
Startup Screen: COMBUSTION (default), JOB MENU, FUEL TYPE
Set which screen you want to see after start countdown is complete.
Auto Hold Display: PUMP STOP (default), NEVER
Select NEVER to continue viewing live measurements while the pump
is stopped.
Combustion Setup: EDIT to select parameters and their slots on the
COMBUSTION screen. To hide a slot completely, assign a parameter to
the slot you want to hide, then assign that parameter to another slot.
Trending Setup: EDIT to select parameters on the TRENDING screen.
SET LIMITS to adjust each parameter’s UPPER (maximum) and LOWER
(minimum) values of the graph. Controlling these limits allows you to
eectively zoom and shift each parameter independently for better
visibility.
Report Contents: EDIT to select what is shown on reports.
Language (status): English (default) This is not adjustable.
Temperature Unit: ° F (default), °C
Pressure Unit: inWC (default), torr, Pa
[%] Reference O2: 6.0% (default) EDIT to adjust the reference oxygen
% used for calculating CO Air Free.
Eciency Calculation: Standard (default), Siegert
Siegerts formula is typically only used in parts of Europe.
Firmware [version]: UPDATE to install new rmware if it becomes
available at www.eldpiece.com