User Manual
01
06
02
07
03
08
04
09
05
10
Quick Start
1 Install six AA batteries into rear battery com-
partment. Batteries included in packaging.
2 Press the center blue button for 1 second to
turn on your new manifold.
3 Connect hoses and pipe clamps to the mani-
fold and the system.
4 See real-time pressure and temperature mea-
surements all at once!
Certifications
C-Tick (N22675)
CE
RoHS Compliant
Description
Your SMAN360 is the top of the line two-
valve Digital Manifold and Micron Gauge for
HVACR professionals. See all your pressures
and temperatures at the same time on the
redesigned large display with bright blue
backlight.
SMAN360 combines high precision, absolute
pressure sensors, a superheat/subcooling
calculator, true micron gauge for vacuum,
and dual temperature measurements. Your
SMAN360 calculates and displays target
superheat and actual superheat to verify proper
charge.
SMAN360 is designed to meet the demands
of HVACR eld service with a rugged rubber
boot for durability, a strong metal hanger for
easy storage and a form tting, water resistant,
padded nylon pouch. SMAN360 has a very
intuitive user interface and long battery life.
Your SMAN comes pre-programmed with
accurate P-T charts for 45 of the most common
refrigerants in the field so you are always
prepared for any job.
Fieldpiece
Digital Manifold &
Micron Gauge
OPERATOR'S MANUAL
Model SMAN360
Controls
1 Insert Type K thermocouple plugs here.
2 Temperature calibration pots.
3 Press to zero atmospheric pressure.
4 Press to calibrate to refrigerant tank.
(See Advanced Pressure Calibration section.)
5 Press to conrm selection.
6 Press/hold to cycle through refrigerants.
7 Press to turn on/o the Hi/Lo vacuum
alarms and hold to change alarm settings.
8 Press to scroll up to adjust values.
9 Press to change units.
10 Press to scroll down to adjust values.
11 Hold 1 second to power on or o. Press to toggle backlight.
12 Press to enter Target Superheat set up mode.
13 Turn clockwise to close High side port.
14 Turn clockwise to close Low side port.
Functions
Superheat and Subcooling
Your SMAN360 can calculate and display both superheat and
subcooling simultaneously.
1 Select the appropriate refrigerant using the REFRIGERANT button.
2 Connect EPA approved refrigerant hoses to low and high side on
SMAN360. Plug Type K thermocouple pipe clamps to SLT and LLT.
3 Connect your SMAN360 to the system:
Superheat: Hand tighten low side hose to suction line service
port. Place the SLT pipe clamp thermocouple on the suction line
between the evaporator and compressor, no closer than 6 inches
to compressor.
Subcooling: Hand tighten high side hose to liquid line service
port. Attach the LLT pipe clamp thermocouple on the liquid line
between the condenser and expansion valve (TXV), as close to the
service port as possible.
4 After turning the system on or making any adjustments to
the system wait 15 minutes before charging by superheat or
subcooling to ensure the system is stabilized.
5 To add or remove refrigerant connect a refrigerant or recovery
tank to the center port on SMAN360. Follow recommened
charging or recovery practices from equipment manufacturer.
Use the low side and high side valves on SMAN360 to charge or
recover refrigerant as needed. Let system stabilize for 15 minutes.
Note: When superheat and/or subcooling cannot be calculated an
"OL" or "-OL" will be displayed. Please check the following:
1 The correct refrigerant is selected on the SMAN.
2 The pipe thermocouples are plugged into SLT/LLT ports and are in
good working condition.
3 The pipe thermocouples are attached in the appropriate location on
the system. See step 3 above for details.
Target Superheat
Target Superheat is useful for charging xed orice air conditioning
systems. Your SMAN will calculate the target superheat for you. Just
manually input the IDWB (indoor wet bulb) and ODDB (outdoor dry
bulb) temperatures into the SMAN.
Manually Input Temperatures
1 Press Target SH button to enter Target SH mode. IDWB will begin
blinking indicating it is ready for an input.
2 Press UP or DOWN ARROW to toggle between IDWB or ODDB
input. Hold ENTER to select which temperature you want to input,
either IDWB or ODDB. The far left digit of IDWB or ODDB will begin
blinking indicating manual input mode is ready.
3 Press the UP or DOWN ARROW to change values and press ENTER
to lock in each digit.
4 Repeat steps 2 and 3 for the other temperature. The calculated
target superheat will show in the center column of the display. A
solid HOLD will display to the left of the TSH calculation indicating
a static target superheat calculation.
Note: If the inputted temperature is out of the calculable range for
IDWB or ODDB an "Err" will ash once and a double beep will
sound. IDWB range (40°F to 125°F, 4.4°C to 51.7°C) and ODDB
range (50°F to 140°F, 10°C to 60°C). Re-input a temperature
within these ranges to calculate target superheat.
Changing Units
Your SMAN can display pressure and temperature measurements in
English, Metric or combination of both units.
1 Press UNITS to enter unit selection screen.
2 Use ARROW to select your desired pressure units. Press ENTER.
3 Use ARROW to select your desired temperature units. Press ENTER
to return to pressure units.
4 Press UNITS to return to normal SMAN display.
Pulling a Vacuum
Follow all manufacturers’ evacuation procedures over those in this
manual.
1 Connect your SMAN360 to your vacuum pump and the system,
then power on your SMAN360.
2 Set up vacuum alarms. These will notify you when you've reached
your desired vacuum and stabilization levels. See Set Vacuum
Alarm instructions below.
3 Pull a vacuum on the system. SMAN360 will automatically sense
the negative pressure and begin to display in inHgV. Once the
vacuum levels are low enough, the display will automatically
change to show vacuum in microns. Once in micron mode the
inHgV readings will no longer display.
4 The rate at which the vacuum levels are changing will be
displayed in microns per minute. The smaller the rate of change,
the closer you are to stabilization.
Set Vacuum Alarms
1 Hold ALARM for 1 second to enter Alarm Set Mode. The rst digit
of LO alarm will blink.
2 Use ARROWS to change the blinking number. Press ENTER to
lock in a digit and move to the next digit. Repeat for all LO alarm
digits.
3 When LO alarm is complete, the rst digit of HI alarm will blink.
Use ARROWS to change the blinking number. Press ENTER to lock
in a digit and move to the next digit. Repeat for all HI alarm digits.
4 When all digits of HI alarm are locked in you will automatically
exit Alarm Set Mode and your target alarm values will be saved.
Note: Anytime while in Alarm Set Mode, you can press ALARM to
toggle between alarm HI set and alarm LO set. Hold ALARM to exit
Alarm Set Mode and save at any time.
Note: "Err" will show if you try to set the HI alarm lower than the LO
alarm, or the LO alarm higher than the HI alarm.
Activate Vacuum Alarms
1 Press ALARM to activate low alarm. Default is 500 microns.
Stopwatch will start. When low alarm target value is reached,
SMAN360 will beep and the stopwatch will restart from zero. You
can monitor how long the vacuum has been under your target.
2 Press ALARM again to deactivate low alarm and activate high
alarm. Default is 1000 microns. When high alarm value is
reached, SMAN360 will beep and the stopwatch will pause. You
can see how long it took to reach your target value.
3 Press ALARM again to deactivate high alarm.
Additional Evacuation Tips to Reach a Deeper Vacuum:
1 Use shortest vacuum rated hoses with largest diameter available.
2 Remove schrader cores and core depressors. Core removal tools
like the "MegaFlow Valve Core Removal Tool" can be purchased
from Appion, Inc. to help with this process.
3 Inspect the rubber seals at both ends of your hoses for damage
that may result in leakage.
4 Do not use hoses with low loss ttings when evacuating or
pulling a vacuum on a system.
5 When the vacuum pump is isolated from the system, a slow rise
in micron levels may signify moisture is still present in the system
but should eventually stablize. A continuous rise in microns levels
to atmosphere may indicate a leak in the hoses, the vacuum
pump connections or the system.
Auto Power O (APO)
To conserve battery life, your SMAN will power down after 30
minutes of inactivity. APO is activated by default and APO displays
above the battery icon. To deactivate, press and hold ENTER while
powering on the SMAN. When deactivated, APO will no longer show
above the battery icon.
LOW HIGH
Pressure (bar)
Pressure (pounds/in
2
)
Pressure (kilopascals or Megapascals)
Negative Pressure (inches of mercury)
Negative Pressure (cm of mercury)
Superheat
Subcooling
Target Superheat
Suction Line Temperature
Liquid Line Temperature
Vapor Saturation Temperature
Liquid Saturation Temperature
Outdoor Dry Bulb
Indoor Wet Bulb
Static Target Superheat Calculation
Hours:Minutes or Minutes:Seconds
Battery Life
Auto Power O Enabled
Selected Refrigerant of System
Vacuum (Microns of Mercury)
Rate of change (Microns per Minute)
Set Mode
High Alarm
Low Alarm
14
11
11
2
2
3 4
5
8
9
10
12
6
7
13
LOW HIGH
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com