| TT10 IN-LINE TURBINE METER
INSTALLATION AND OPERATION MANUAL
Reset Accumulated Total
1. Press and hold buttons 2 and 4 for three seconds. Current firmware version will be displayed.
2. Press the button to display the accumulated total. If over 9999 units, the accumulated total will scroll across the display.
3. Hold buttons and 4 together for five seconds to reset the accumulated total.
4. Press the 2 button twice to return to normal operating mode.
Battery Replacement
Batteries should be replaced when the LOW battery icon begins to blink
on the display. The meter totals, configuration, and calibration are
preserved while the batteries are being replaced, so there is no need
to recalibrate afterwards.
The batteries are accessed by removing the display as described in the
section showing how to rotate the display. The Battery Foam may be stuck
to the batteries. Carefully remove it and reinstall at assembly. The Battery
Foam protects the batteries and prevents them from becoming dislodged if
the meter is handled roughly.
Pay attention to the polarity of the batteries as marked inside the battery
holder. The meter will not operate if the batteries are installed incorrectly
and damage may occur.
When replacing batteries, make the area safe of explosive vapors or move the meter electronics to a safe area before replacing them. To reduce
the risk of explosion, DO NOT mix old batteries with new batteries or mix batteries of different manufacturers.
Batteries may ONLY be replaced with the following:
• Duracell MN1500 • Duracell QU1500 • Energizer E91
WARNING
FOAM PAD
AA BATTERIES
K-Factor Check
You can check the internal calibration factor for the meter. The displayed value is the number of sensor counts per selected unit of measure and
CAL/Fluid factor. The factory may ask you for this value if troubleshooting the Meter.
1. Hold buttons 3 and 4 for three seconds. The Meter will display "FLSH".
2. Press and hold buttons and 3 together. The number of counts per unit of measure will be displayed.
3. Press button 2 to return to normal operating mode.
Maintenance and Storage
If the meter will be stored for an extended period of time without use and is being used with a fluid that may leave a residue, it should be flushed with
a compatible fluid to prevent sticking and binding of the turbine assembly.
The exterior of the meter may be wiped down with a mild water based detergent and damp cloth to clean the display. Do not use a power washer to
clean the meter.
Use of compressed air WILL cause permanent damage to the meter. Always use a gentle stream of compatible fluid to rinse the interior of the meter.
NOTICE
Flush Mode (FLSH)
This mode allows the user to flush the meter without adding volume to the totalizator. It is useful for cleaning or switching between different liquids to
maintain measurement accuracy.
1. Hold buttons 3 and 4 for three seconds. Meter will display "FLSH".
2. Press button 2 to return to normal operating mode.