6.1.2 DRYING:
6.1.2.a After washing and rinsing all the com-
ponents, shake them to remove excess water
and reassemble them as described in the fol-
lowing paragraph DEVICE REASSEMBLY.
At this point, to eliminate the water deposited
in the nebulisation and mesh cup (2), press key
1A (gure opposite) and wait for the device to
stop nebulising. Then switch o the unit.
WARNING This operation is very important be-
cause it avoids limescale deposit in the mesh
microholes, which could compromise the cor-
rect nebulisation in the subsequent therapeu-
tic application.
6.1.2.b To complete the drying of all compo-
nents, leave them in the open air and in a dry
place (e.g. not in the bathroom) as shown in the
gure opposite.
WARNING
Do not immerse the Device Body (1) in water
and do not wash it under running tap water.
6.2 CLEANING Device Body
Use a cloth lightly moistened with antibacterial
detergent to clean the Device Body (1).
Then dry it with a soft paper towel.