User guide
C.3 COMMANDS SUMMARY
Manual DAS 72.1 Page 27
Command Short Description Usage Parameter Values Typical Answer
AA
Get/set analog action Analog action 0 through 8 A+00000/OK/ERR
AD
Network address Read/modify Network address 0 through 255 A:014/OK/ERR
AG
Absolute gain calibrate (TAC protected) Cailbrate gain using absolute sensitivity in mV/V (without dp) +/-32000; 0 through 99999 G+2.0000/OK/ERR
AH
Get/set analog high Analog calibration point 20 mA (high) -99999 through 9999 H+10000/OK/ERR
AIn
Action input n Function of logic input 1, 2 or 3 0 through 14 In:+00006
AL
Get/set analog low Analog calibration point 4 mA (low) -99999 through 99999 L+00000/OK/ERR
An
Get/set setpoint n action Setpoint action 1, 2 or 3 0 through 8 An:+00000/OK/ERR
AS
Save analog output parameters Save AL, AH, AA to EEPROM none OK/ERR
AZ
Absolute zero calibrate (TAC protected) Calbrate zero using absolute zero in mV/V (without decimal point) -32000 through 32000 Z+0.000/OK/ERR
BR
Baudrate Read/modify baudrate setting 9600 through 115200 B 9600
CE
Calibrate enable Open legal parameters (CE 'TAC') 0 through 65535 E+00001/OK/ERR
CG
Calibrate gain (TAC protected) Calibrate gain at some load >> zero 0 through 99999 G+20000/OK/ERR
CI
Calibrate min (TAC protected) Read/modify min 0 through -99999 I+09000/OK/ERR
CL
Close all connections Close all devices (OP XXX implies CL) none none
CM
Calibrate max (TAC protected) Read/modify max 0 through 99999 M+30000/OK/ERR
CS
Calibrate save (TAC protected) Save CM, CZ, AZ, CG, AG, DS, DP, ZT to EEPROM none OK/ER
CZ
Calibrate zero (TAC protected) Calibrate zero - no load on platform none OK/ER
DP
Decimal point (TAC protected) Read/modify decimal point position 0 through 5 P+00005/OK/ERR
DS
Display step size (TAC protected) Read/modify the display step size 1, 2.5, 10, 20, 50, 100, 200, 500 S+00001/OK/ERR
DX
Duplex Select half (0) or full (1) duplex 0 oder 1 X+000/OK/ERR
FD
Factory default (TAC protected) Load factory defaults into EEPROM none OK/ER
FL
Filter setting Read/modify filter setting 0 through 8 F+00008/OK/ERR
FM
Filter mode Read/modify filter mode setting 0 oder 1 M+00001/OK/ERR
GA
Get Average weight Returns the current average value none A+00000
GG
Get Gross value Returns the gross value none G+01100
GH
Get Hold value Returns the hold value none H+01000/OK/ERR
GM
Get Peak value (maximum) Returns the peak value none M+00100
GN
Get Net value Returns the net value none N+00000
GO
Get Peak to Peak value Returns the Peak to Peak value none O+01000/OK/ERR
GS
Get sample Returns the A/D value none S+000000
GT
Get Tare value Returns the tare value none T+00000
GV
Get Valley value (minimum) Returns the valley value none V+01000/OK/ERR
GW
Get long Weight information Get long weight information none W+00100+01100010F
Hn
Get/set setpoint n hysteresis Setpoint hysteresis 1, 2 or 3 -99999 through 99999 Hn:+00000/OK/ERR