FLINTEC Remote Display TYP FRD-57 Flintec GmbH Bemannsbruch 9 74909 Meckesheim Germany 9 In case of any questions… Tel. : +49 6226 92400 9 You can visit us http://www.flintec.net Remote Display FRD-57 Datum Code 27.09.
FLINTEC Connection FRD- 57 The connection should be conducted with an 2-wire shielded cable. The shield should be connected at the FT-Indicator . Fig.1 shows the terminal block, Jumper and Dip switch at the remote indicator. Fig. 1 Spannungsversorgung 10VAC A RS485 B JP1 +Tx JP2 20mA current loop RS232 1 2 3 4 -Tx Tx SGND S1 SHLD ST2 Settings of the Dip Switch ( S1 ) Fig.
FLINTEC Sample for Settings Indikator FT-01 / FT-02 . Connection via RS-232 Interface Connect the Tx and GND at the FT-Indicator with Tx and SGND at the remote Indicator. The shield should conneted via the screwed cable gland (single-edge) Please consider the max. cable length with RS-232 respectively TTY Interface is 15 meters and with RS-485 max. 1000 meters should not exceed. The Jumper JP1 and JP2 at the remote indicator must be open .