8.7. Synchronized Output Commands – HW, GH
For systems requiring simultaneous sampling of data two commands are available:
8.7.1. HW Hold Weight
Issuing the HW command, which has no parameters, will latch the current net weight in a register for later
readout in all LDU’s in a common network regardless of their individual address.
Note: The LDU’s will not respond to the HW command on the interface lines – it would lead to bus contention
Master (PC / SPS) sends Slave (LDU XX.X) responds Meaning
HW No response
Hold weight value in register
8.7.2. GH Get Hold Weight
Issuing the GH command, which has no parameters, will return the current hold weight in the format N+000100
from the currently addressed LDU.
Master (PC / SPS) sends Slave (LDU XX.X) responds Meaning
OP 1 OK
Open device #1
GH H+01.100
Hold weight value: 1,100 d
All other channels (LDU with other address) have to be requested one by one.