User's Manual
Table Of Contents
- 1 Disclaimers
- 2 Safety information
- 3 Notice to user
- 4 Customer help
- 5 Quick Start Guide
- 6 List of accessories and services
- 7 Description
- 8 Operation
- 8.1 Charging the battery
- 8.2 Turning on and turning off the camera
- 8.3 Saving an image
- 8.4 Recalling an image
- 8.5 Deleting an image
- 8.6 Deleting all images
- 8.7 Measuring a temperature using a spotmeter
- 8.8 Measuring the hottest temperature within an area
- 8.9 Measuring the coldest temperature within an area
- 8.10 Hiding measurement tools
- 8.11 Changing the color palette
- 8.12 Working with color alarms
- 8.13 Changing image mode
- 8.14 Changing the temperature scale mode
- 8.15 Setting the emissivity as a surface property
- 8.16 Setting the emissivity as a custom material
- 8.17 Changing the emissivity as a custom value
- 8.18 Changing the reflected apparent temperature
- 8.19 Changing the distance between the object and the camera
- 8.20 Performing a non-uniformity correction (NUC)
- 8.21 Configuring Wi-Fi
- 8.22 Changing the settings
- 8.23 Updating the camera
- 9 Technical data
- 10 Mechanical drawings
- 11 CE Declaration of conformity
- 12 Cleaning the camera
- 13 Application examples
- 14 About FLIR Systems
- 15 Definitions and laws
- 16 Thermographic measurement techniques
- 17 History of infrared technology
- 18 Theory of thermography
- 19 The measurement formula
- 20 Emissivity tables
Thermographic measurement techniques16
3. Measure the radiation intensity (= apparent temperature) from the reflection source
using the following settings:
• Emissivity: 1.0
• D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
Figure 16.3 1 = Reflection source Figure 16.4 1 = Reflection source
You can not use a thermocouple to measure reflected apparent temperature, because a
thermocouple measures temperature, but apparent temperatrure is radiation intensity.
16.2.1.1.2 Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
#T559828; r. AK/40423/40423; en-US