User Manual
Table Of Contents
- 1 Legal disclaimer
- 2 Warnings & Cautions
- 3 Notice to user
- 4 Customer help
- 5 Quick Start Guide
- 6 Parts lists
- 7 A note about ergonomics
- 8 Camera parts
- 9 Screen elements
- 10 Navigating the menu system
- 11 External devices and storage media
- 12 Pairing Bluetooth devices
- 13 Configuring Wi-Fi
- 14 Fetching data from external Extech meters
- 15 Handling the camera
- 15.1 Charging the battery
- 15.2 Inserting the battery
- 15.3 Removing the battery
- 15.4 Turning on and turning off the camera
- 15.5 Adjusting the angle of lens
- 15.6 Mounting an additional lens
- 15.7 Removing an additional infrared lens
- 15.8 Attaching the sunshield
- 15.9 Using the laser pointer
- 15.10 Calibrating the compass
- 15.11 Calibrating the touchscreen LCD
- 16 Working with images and folders
- 17 Working with fusion
- 18 Working with video
- 19 Working with measurement tools and isotherms
- 20 Annotating images
- 21 Programming the camera
- 22 Changing settings
- 23 Cleaning the camera
- 24 Technical data
- 25 Pin configurations
- 26 Dimensions
- 27 Application examples
- 28 About Flir Systems
- 29 Glossary
- 30 Thermographic measurement techniques
- 31 History of infrared technology
- 32 Theory of thermography
- 33 The measurement formula
- 34 Emissivity tables
Working with measurement tools and isotherms19
19.6 Moving measurement tools
19.6.1 Procedure
Follow this procedure:
1. Push the Menu/Back button.
2. On the main menu, go to the Tools button
and push the joystick.
3. On the Tools menu, select Adjust tools
and push the joystick.
4. Select the measurement tool that you wish to move. This will display a submenu.
5. On the submenu, select Move and push the joystick. This will make the center of the
measurement tool turn blue. You can now move the measurement tool using the
joystick.
19.7 Resizing areas
19.7.1 Procedure
Follow this procedure:
1. Push the Menu/Back button.
2. On the main menu, go to the Tools button
and push the joystick.
3. On the Tools menu, select Adjust tools
and push the joystick.
4. Select the measurement tool that you wish to resize. This will display a submenu.
5. On the submenu, select Resize and push the joystick. This will create resizing han-
dles for the area. You can now resize the area using the joystick.
19.8 Changing object parameters
19.8.1 General
For accurate measurements, you must set the object parameters. This procedure de-
scribes how to change the parameters.
19.8.2 Types of parameters
The camera can use these object parameters:
• Emissivity, which determines how much of the radiation originates from the object as
opposed to being reflected by it.
• Reflected apparent temperature, which is used when compensating for the radiation
from the surroundings reflected by the object into the camera. This property of the ob-
ject is called reflectivity.
• Object distance, i.e. the distance between the camera and the object of interest.
• Atmospheric temperature, i.e. the temperature of the air between the camera and the
object of interest.
• Relative humidity, i.e. the relative humidity of the air between the camera and the object
of interest.
• External optics temperature, i.e., the temperature of any protective windows etc. that
are set up between the camera and the object of interest. If no protective window or
protective shield is used, this value is irrelevant.
• External optics transmission, i.e., the optical transmission of any protective windows,
etc. that are set up between the camera and the object of interest.
19.8.3 Recommended values
If you are unsure about the values, the following are recommended: