User Manual
Table Of Contents
- 1 Legal disclaimer
- 2 Warnings & Cautions
- 3 Notice to user
- 4 Customer help
- 5 Quick Start Guide
- 6 Parts lists
- 7 A note about ergonomics
- 8 Camera parts
- 9 Screen elements
- 10 Navigating the menu system
- 11 External devices and storage media
- 12 Pairing Bluetooth devices
- 13 Configuring Wi-Fi
- 14 Fetching data from external Extech meters
- 15 Handling the camera
- 15.1 Charging the battery
- 15.2 Inserting the battery
- 15.3 Removing the battery
- 15.4 Turning on and turning off the camera
- 15.5 Adjusting the angle of lens
- 15.6 Mounting an additional lens
- 15.7 Removing an additional infrared lens
- 15.8 Attaching the sunshield
- 15.9 Using the laser pointer
- 15.10 Calibrating the compass
- 15.11 Calibrating the touchscreen LCD
- 16 Working with images and folders
- 17 Working with fusion
- 18 Working with video
- 19 Working with measurement tools and isotherms
- 20 Annotating images
- 21 Programming the camera
- 22 Changing settings
- 23 Cleaning the camera
- 24 Technical data
- 25 Pin configurations
- 26 Dimensions
- 27 Application examples
- 28 About Flir Systems
- 29 Glossary
- 30 Thermographic measurement techniques
- 31 History of infrared technology
- 32 Theory of thermography
- 33 The measurement formula
- 34 Emissivity tables
Thermographic measurement techniques30
3. Measure the radiation intensity (= apparent temperature) from the reflecting source
using the following settings:
• Emissivity: 1.0
• D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
1 = Reflection source
Note
Using a thermocouple to measure reflected apparent temperature is not recommended for two impor-
tant reasons:
• A thermocouple does not measure radiation intensity
• A thermocouple requires a very good thermal contact to the surface, usually by gluing and covering
the sensor by a thermal isolator.
30.2.1.1.2 Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
5. Measure the apparent temperature of the aluminum foil and write it down.
Measuring the apparent temperature of the aluminum foil.
#T559772; r.5948/5948; en-US