User's Manual
Table Of Contents
- 1 Disclaimers
- 2 Safety information
- 3 Notice to user
- 4 Customer help
- 5 Quick start guide
- 6 Register the camera
- 7 A note about ergonomics
- 8 Camera parts
- 9 Screen elements
- 10 Navigating the menu system
- 11 Handling the camera
- 11.1 Charging the battery
- 11.2 Installing and removing the camera battery
- 11.3 Turning on and turning off the camera
- 11.4 Adjusting the angle of lens
- 11.5 Adjusting the infrared camera focus manually
- 11.6 Autofocusing the infrared camera
- 11.7 Continuous autofocus
- 11.8 Operating the laser distance meter
- 11.9 Measuring areas
- 11.10 Connecting external devices and storage media
- 11.11 Moving files to a computer
- 11.12 Assigning functions to the programmable buttons
- 11.13 Using the camera lamp as a flash
- 11.14 Changing camera lenses
- 11.15 Neck strap
- 11.16 Hand strap
- 12 Saving and working with images
- 13 Working with the image archive
- 14 Achieving a good image
- 15 Working with image modes
- 16 Working with measurement tools
- 17 Working with color alarms and isotherms
- 18 Annotating images
- 19 Programming the camera (time-lapse)
- 20 Recording video clips
- 21 Screening alarm
- 22 Pairing Bluetooth devices
- 23 Configuring Wi-Fi
- 24 Fetching data from external FLIR meters
- 25 Changing settings
- 26 Cleaning the camera
- 27 Technical data
- 27.1 Online field-of-view calculator
- 27.2 Note about technical data
- 27.3 Note about authoritative versions
- 27.4 FLIR T530 24°
- 27.5 FLIR T530 42°
- 27.6 FLIR T530 24° + 14°
- 27.7 FLIR T530 24° + 42°
- 27.8 FLIR T530 24° + 14° & 42°
- 27.9 FLIR T530 42° + 14°
- 27.10 FLIR T540 24°
- 27.11 FLIR T540 42°
- 27.12 FLIR T540 24° + 14°
- 27.13 FLIR T540 24° + 42°
- 27.14 FLIR T540 24° + 14° & 42°
- 27.15 FLIR T540 42° + 14°
- 28 Mechanical drawings
- 29 Application examples
- 30 About FLIR Systems
- 31 Terms, laws, and definitions
- 32 Thermographic measurement techniques
- 33 The secret to a good thermal image
- 34 About calibration
- 34.1 Introduction
- 34.2 Definition—what is calibration?
- 34.3 Camera calibration at FLIR Systems
- 34.4 The differences between a calibration performed by a user and that performed directly at FLIR Systems
- 34.5 Calibration, verification and adjustment
- 34.6 Non-uniformity correction
- 34.7 Thermal image adjustment (thermal tuning)
- 35 History of infrared technology
- 36 Theory of thermography
- 37 The measurement formula
- 38 Emissivity tables
The measurement formula
37
magnitudes of the three radiation terms. This will give indications about when it is impor-
tant to use correct values of which parameters.
The figures below illustrates the relative magnitudes of the three radiation contributions
for three different object temperatures, two emittances, and two spectral ranges: SW and
LW. Remaining parameters have the following fixed values:
• τ = 0.88
• T
refl
= +20°C (+68°F)
• T
atm
= +20°C (+68°F)
It is obvious that measurement of low object temperatures are more critical than measur-
ing high temperatures since the ‘disturbing’ radiation sources are relatively much stron-
ger in the first case. Should also the object emittance be low, the situation would be still
more difficult.
We have finally to answer a question about the importance of being allowed to use the
calibration curve above the highest calibration point, what we call extrapolation. Imagine
that we in a certain case measure U
tot
= 4.5 volts. The highest calibration point for the
camera was in the order of 4.1 volts, a value unknown to the operator. Thus, even if the
object happened to be a blackbody, i.e. U
obj
= U
tot
, we are actually performing extrapola-
tion of the calibration curve when converting 4.5 volts into temperature.
Let us now assume that the object is not black, it has an emittance of 0.75, and the trans-
mittance is 0.92. We also assume that the two second terms of Equation 4 amount to 0.5
volts together. Computation of U
obj
by means of Equation 4 then results in U
obj
= 4.5 /
0.75 / 0.92 – 0.5 = 6.0. This is a rather extreme extrapolation, particularly when consider-
ing that the video amplifier might limit the output to 5 volts! Note, though, that the applica-
tion of the calibration curve is a theoretical procedure where no electronic or other
limitations exist. We trust that if there had been no signal limitations in the camera, and if
it had been calibrated far beyond 5 volts, the resulting curve would have been very much
the same as our real curve extrapolated beyond 4.1 volts, provided the calibration algo-
rithm is based on radiation physics, like the FLIR Systems algorithm. Of course there
must be a limit to such extrapolations.
Figure 37.2 Relative magnitudes of radiation sources under varying measurement conditions (SW cam-
era). 1: Object temperature; 2: Emittance; Obj: Object radiation; Refl: Reflected radiation; Atm: atmos-
phere radiation. Fixed parameters: τ = 0.88; T
refl
= 20°C (+68°F); T
atm
= 20°C (+68°F).
#T810253; r. AA/42549/42549; en-US