Table7.2Current(TRMS)
Function Ran
g
e Accurac
y
DCA
60.00 A
±(2%+5dgt)
600.0 A
ACA
60.00 A
±(2%+5dgt)

45~400Hz
600.0 A
ACAVFD
60.00 A
±(2%+5dgt)45~65Hz
±(6%+5dgt)65~400Hz
600.0A
Notes:
Overloadprotection:600A(rms)
PositionError:±1%ofreading.
ACConversionTypean d additionalaccuracyissameasACVoltage.
DCAaffectedbythetemp er at ur e andresidualmagnetism;useDCAZerofuncti o n tocompensate.
Ifthemetermeasuresa4000countsignalandtheCrestFactorofthesignalismorethen3.0,the
readingmaynotmeetspecifiedtolerances.Fornonsinewaves(50/60Hz),addthefollowingCrest
Factorcorrections:
Fornonsinewaves(50/60Hz),addthefollowingcrestfactorcorrections:
Foracrestfactorof1.0–2.0,add3.0%totheaccuracy.
Foracrestfactorof2.0–2.5,add5.0%totheaccuracy.
Foracrestfactorof2.5–3.0,add7.0%totheaccuracy.
Table7.3Frequency(ACAandACV)
Function Range
Accuracy
Frequency
10.0~600.0 Hz
±(0.1%+2dgt)
6.000 kHz
60.00 kHz
Notes:
Overloadprotection:1000V(rms)and600A(rms)
TriggerSensitivity:
>5V(rms)forACV10Hz~10kH z
>15V(rms)forACV10kHz~60kH z
>4A(rms)forACA10Hz ~1kHz
>8A(rms)forACA1kHz~10kH z
Readingwillbe0.0
forsign al sbelow10.0Hz