USER MANUAL FLIR MODEL CM55 and CM57 FLEX CLAMP METERS with Bluetooth®
Table of Contents 1. 2. 3. 4. DISCLAIMERS 3 1.1 Copyright 3 1.2 Quality Assurance 3 1.3 Documentation 3 1.4 Disposal of Electronic Waste 3 SAFETY 4 2.1 FCC Compliance 6 2.2 Industry Canada Compliance 7 INTRODUCTION 8 3.1 Key Features 8 DESCRIPTIONS 9 4.1 Meter Description 5. 9 4.2 Display Icon Description 10 4.3 Control buttons 10 OPERATION 11 5.1 Powering the Meter 6. 7. 8. 9. 11 5.2 Work Lights 11 5.3 Data Hold 11 5.4 AC Current Measurements 12 5.
1. Disclaimers 1.1 Copyright © 2015, FLIR Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual or otherwise, without the prior written permission of FLIR Systems.
2. Safety Safety Notes Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes. FLIR Systems reserves the right to discontinue models, parts or accessories, and other items, or to change specifications at any time without prior notice. Remove the batteries if the device is not used for an extended period of time. Warning Statements WARNINGS identify hazardous conditions and actions that could cause BODILY HARM or DEATH.
Cautions Do not use the device for a procedure that it is not made for. This can cause damage to the protection. This symbol, adjacent to another symbol, indicates the user must refer to the manual for further information. Do not apply or remove clamp from HAZARDOUS LIVE conductors Equipment protected by double or reinforced insulation Battery symbol Conforms to EU directives Do not discard this product in household trash.
2.1 FCC Compliance This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: 1. 2. This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
2.2 Industry Canada Compliance This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: This device complies with Industry Canada license‐exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. CAUTION Exposure to Radio Frequency Radiation.
3. Introduction Thank you for choosing the FLIR Flex Clamp with Bluetooth® that can measure up to 3000A AC rms. The CM57 is the 18” (45.7cm) clamp version and the CM55 is the 10” (25.4cm) clamp version, otherwise both meters are the same. These devices are professional CAT IV 600V CAT III 1000V instruments that offer Data Recording, Bluetooth®, Auto Power OFF, Data Hold, Display backlight, and high power Work Light features.
4. Descriptions 4.1 Meter Description 1. Work Lights 2. Flexible Current Clamp Coil 3. Display 4. INRUSH / REC (Record) button 5. Power / Work Light button 6. RANGE / SEND button 7. HOLD / Start‐Stop Record / Bluetooth® button 8. Clamp Lock mechanism Note that battery compartment is located on the back of the meter 2 8 1 3 4 7 5 6 Fig.
4.2 Display Icon Description Bluetooth® Display Hold Amperes Battery status Inrush Current Auto Power Off Data Transmission Memory Mode Auto Range 4.
5. Operation Note: Before operating the device, please read and understand all Warning and Caution statements and follow all instructions and notes. 5.1 Powering the Meter The meter is powered by two (2) AAA 1.5V batteries (located in the compartment on the back of for > 2 seconds to switch the device ON or OFF. the meter). Press and hold the Power button 5.1.1 Auto Power OFF (APO) The meter switches OFF automatically after an approximately 10‐minute period of inactivity.
5.4 AC Current Measurements WARNING: Ensure that power to the device under test is OFF before starting this procedure. Switch power to the device under test ON only after the clamp has been safely attached to the device under test. CAUTION: Do not move fingers above the LCD at any time during a test. 1. 2. 3. 4. 5. 6. 7. 8. Switch the meter OFF and switch power to the device under test OFF. Turn the clamp lock (1) counter‐clockwise to release the flexible clamp (2).
5.4.1 Inrush Current mode The meter has the ability to capture an inrush current signal using a 100ms sampling window. The sampling window opens only when the threshold current (see below) is detected. When detecting an input current ±50 digits of the selected range the meter will calculate the RMS values for a 100mS period and display this value. Refer to illustration below. For the 30A range the minimum triggering current threshold is 0.
5.5 Data Recording and Bluetooth® Data Transfer 5.5.1 Data Recording When prompted, this meter can store readings into its internal memory automatically for later transfer over Bluetooth® (real‐time data streaming over Bluetooth® is also possible as explained below) using the FLIR TOOLS application. The maximum number of records that can be stored is 20,000 and the fixed sampling (recording) interval is 1 minute. 1. Press the REC button >2 seconds to enter the Data Record Mode.
5.5.3 Transferring Recorded Readings with Bluetooth® To transfer bulk logged readings from the meter’s internal memory to a Bluetooth® paired device running the FLIR TOOLS application please read the steps below. Note that the Bluetooth range is 32’ (10m) maximum. 1. Press and hold the Bluetooth button 2. 3. 4. Press and hold the SEND button until the percent (%) symbol appears The data is now being transmitted. The percentage shown is the percentage of data transmitted.
6. Maintenance 6.1 Cleaning and Storage Clean the meter with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for an extended period, remove the batteries and store them separately. 6.2 Battery Replacement CAUTION: Remove the meter from the conductor under test and switch the meter OFF before opening the battery compartment. 1. With a Phillips screwdriver, remove the battery compartment screw on the back of the meter. 2.
7. Specifications 7.1 General specifications Clamp Jaw Flexible type with locking mechanism Coil diameter 0.3” (7.5mm); Coil tip (item 2 in Fig. 5‐1): 0.5” (13mm) Coil Bend Radius 3.1” (80mm) for CM57; 1.5” (38mm) for CM55 Display 3000 count LCD with backlight and multi‐function indicators Display update rate 2 times per second Low Battery indication Battery symbol appears empty and flashing Over‐range indication ‘OL’ display Work Lights Two White LEDs Measurement rate 1.
7.2 AC Current Electrical specifications Function AC Current Range Overload Reading (OL) Resolution Accuracy (of reading) 45 to 500Hz 30.00 A AC 33.00 A AC 0.01A ±(3.0% + 5 digits) 300.0 A AC 330.0 A AC 0.1A ±(3.0% + 5 digits) 3000 A AC 3300 A AC 1A ±(3.0% + 5 digits) Notes: Accuracy is given as ± (% of reading + least significant counts) at 73.4F ±9F (23C ±5C) with relative humidity lower than 80%. Accuracy is specified for a period of one year after calibration.
8. Technical Support Main Website http://www.flir.com/test Technical Support Website http://support.flir.com Technical support Email TMSupport@flir.com Service/Repair Support Email Repair@flir.com Support Telephone number +1 855‐499‐3662 (toll‐free) 9. Warranty 9.1 FLIR Global Limited Lifetime Warranty A qualifying FLIR Test and Measurement product (the “Product”) purchased either directly from FLIR Commercial Systems Inc.
MANUALS, FUSES, OR DISPOSABLE BATTERIES. FLIR FURTHER EXPRESSLY DISCLAIMS ANY WARRANTY COVERAGE WHERE THE ALLEGED NONCONFORMITY IS DUE TO NORMAL WEAR AND TEAR, OTHER ALTERATION, MODIFICATION, REPAIR, ATTEMPTED REPAIR, IMPROPER USE, IMPROPER MAINTENANCE, NEGLECT, ABUSE, IMPROPER STORAGE, FAILURE TO FOLLOW ANY PRODUCT INSTRUCTIONS, DAMAGE (WHETHER CAUSED BY ACCIDENT OR OTHERWISE), OR ANY OTHER IMPROPER CARE OR HANDING OF THE PRODUCTS CAUSED BY ANYONE OTHER THAN FLIR OR FLIR’S EXPRESSLY AUTHORIZED DESIGNEE.
Corporate Headquarters FLIR Systems, Inc. 2770 SW Parkway Avenue Wilsonville, OR 97070 USA Telephone: +1 503‐498‐3547 Customer Support Technical Support Website Technical Support Email Service and Repair Email Customer Support Telephone http://support.flir.com TMSupport@flir.com Repair@flir.com +1 855‐499‐3662 (toll free) Publication Identification No.