5.12NormalHoldModeandAutoHoldMode
Themeterhastwotypesofholdmodes:normalholdmodeandautoholdmode.
5.12.1NormalHoldMode
In Normal hold mode, the meter freezes and displays the last reading from the main display and
continues to display this value.
To enter/exit Normal hold mode, press the
button. In Hold mode, the indicator is
displayed.
5.12.2AutoHoldMode
In Auto hold mode, the secondary display freezes the last reading from the main display and
continues to display this value. The current reading is displayed on the main display. The held
reading (on the secondary display) will not change unless the difference between this held
reading and any new reading is greater than 50 digits.
Auto hold limit:
Function switch in V position: < 0.1 V
Function switch in LoZ position: < 0.1 V
Function switch in mV position: < 1 mV
Function switch in other positions: no limit.
To enter/exit Auto hold mode, press the
button. In Auto hold mode, the indicator is
displayed and flashing.
5.13LockedMode
InLockedmodethemeterignoresallbuttonpressesexcept .Theautopowerofffunction,
seesection5.1.1AutoPoweroff,isdisabledinLockedmode.
Pressandholdthe
buttonfor3secondstoenter/exitLockedmode.InLockedmode,the
indicatorisdisplayed.