Product Manual
FLIRDM92USERGUIDEDocumentIdentifier:DM92‐en‐US_AB
22
7. Specifications
7.1Generalspecifications
Maximumvoltage: 1000VDCor1000VACRMS.
DisplayCount: 4000/40000.
PolarityIndication: Automatic,positiveimplied,negativeindicated.
Over‐rangeIndication: OL
MeasuringRate: 10samplespersecond
PowerRequirements: 6x1.5VAAAalkalinebatteries
BatteryLife: Approx.100hourswithbatteries(backlightand
worklightoff).
LowBatteryVoltage: Approx.7.0V.
AutoPowerOff: Default10minutes
OperatingTemp/RH: 14°Fto86°F(‐10°Cto30°C),<85%RH.
86°Fto104°F(30°Cto40°C),<75%RH.
104°Fto122°F(40°Cto50°C),<45%RH.
StorageTemperature/RH:‐22°Fto140°F(‐
30°Cto‐60°C),0‐80%RH(batteriesnotfitted).
TemperatureCoefficient: 0.1x(specifiedaccuracy)/°C,<64.4°F(18°C),>82.4°F(28°C).
OperatingAltitude: 6560’(2000m).
CalibrationCycle: Onceperyear.
Weight: 16.4oz.(465g),includingbatteries.
Dimensions: 2.0x3.2x7.4in(52x83x188mm),withholster.
Safety:
ComplieswithIEC61010‐1CATIV‐600V,CATIII‐1000V,IEC61010‐2‐033.
CAT ApplicationField
I Circuitsnotconnectedtomains.
II Circuitsdirectlyconnectedtoalow‐voltage installation.
III Buildinginstallation.
IV Sourceofthelow‐voltageinstallation.
EMC:EN61326‐1.
Pollutiondegree:2.
Dropprotection:6.6’(2m).