Product Manual

FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
15
5.5.2AutoHoldmode
IfAutoHoldmodeisswitchedONinthesettingsmenu,presstheHOLDbuttontoactivate
AutoHoldmode(the‘H’displayiconwillappearflashing).InAutoHoldmodereadingsare
capturedinthesaywayasinstandardHoldmode.Thedifferenceisthatthecaptured
reading,displayed
onthesmaller,secondarydisplaydigits,willupdatewheneverareading50
unitshigherorlowerthanthecapturedreadingismeasured.ToexittheAutoHoldmode
presstheHOLDbutton.
5.6VoltageMeasurements
1.Setthefunctionswitchtothe position.
2.InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3.UsetheMODEbuttontoselectACorDCvoltagemeasurements.
The indicatorwillbedisplayedforACmeasurements.
The
indicatorwillbedisplayedforDCmeasurements.
4.Connecttheprobeleadsinparalleltothepartundertest
5.Readthevoltagevalueonthedisplay
5.7InsulationResistanceMeasurements
1. Usetherotaryswitchtoselecttheinsulationvoltage(50V,100V,250V,500V,or1kV).
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheInsulation
terminal.
3. SelectINSULtoenabletheInsulationResistanceFunctionmodeasdescribedinsection5.12.1
Selectingmodes.
4. DuringtheTest,theinsulationvoltagewillappearinthesecondary(upper)readingarea.The
resistancewillbeshowninthemainreadingarea.
5. PressTESTOKtostarttest.
6. Tolockthetestforcontinuoustesting,presstheHOLD/LOCKbutton(lockiconwillappear)and
thenpressTESTOKtotestcontinuously.
7. TheTESTiconontheLCDwillbeshownwhilethevoltageisoutputted.
8. PressTESTOKagaintostopthetest
NOTE:The99pointDataRecordmemoryfunctioncannotbeactivatedintheusualwaywhile
usingthe
InsulationResistancefunction(sincetheTESTOKbuttonisusedfortestinginthis
caseandnotforsavingreadings).Thelastreading,however,isheldonthedisplayandcanthen
besavedinthenormalwayasdescribedintheExtendedFunctionalitysectionbelowunder
Section5.12.8DataRecordMode.
5.7.1PI/DARTest
Note:PI(PolarizationIndex);DAR(DielectricAbsorptionRatio)
1. SelectthePIorDARtestasdescribedinsection5.12.1Selectingmodes.
2. PresstheTESTOKbuttontotest.
3. PresstheMODEbuttontoshowthetesttimeattheupperreadingarea.(LCDwillshow
CLOCKiconandtheelapsedtime).PIisatenminutetestandDARisaoneminutetest.
4. Thetestwillstopautomaticallyandtheresult(ratio)willbeshownonthemaindisplay
area.ThePASS/FAILLEDwillindicatethetestresult(GreenforPASS,andRedforFAIL).