7. Specifications
7.1Generalspecifications
Display 4,000countwithbargraph
Controls
9positionrotaryswitch
4waySelectorPadwithcenterOK/TESTbutton
LED:Pass(Green),Fail(Red)
(5)dedicatedfunctionbuttons:Mode,Cancel,Hold/Lock,
Bluetooth,Backlight
Backlight WhiteLED
WorkLights WhiteLEDarray
MeasurementRate 2samplespersecond,nominal
DisplayRate 2timespersecond
InputImpedance 3MΩVDC/VAC<100pF,
ACVoltageBandwidth 50Hz500Hz
PowerSupply 6x‘AAA’(LR03)batteries
BatteryLifeConsiderations Themetercanperform1200EarthBondresistance
measurementswithnewalkalinebatteries.Theseare
standardtestsof1Ωwithadutycycleof5secondsONand
25secondsOFF.
InsulationResistance:Themetercanperform300Insulation
Resistancetestswithnewalkalinebatteries.Theseare
standard
testsof1MΩwithadutycycleof5secondsONand
25secondsOFF.ThisdoesnotincludeBluetoothorBacklight
use.
AutoPowerOff(APO)
UserprogrammableOFFto30minutesin1minute
increments,withaudibleprealert(20seconds)
MeasurementTypes TrueRMSACVolts,DCVolts,EarthBondResistance,
InsulationResistance,Capacitance,Diode,andContinuity
OvercurrentProtection(fuse) 440mA/1000V,ceramicfastblowfusewithaminimum
InterruptRatingof10kA
PolarizationIndexandDielectric
AbsorptionRatioTests
PI(PolarizationIndex)=R10min/R1min
DAR(DielectricAbsorptionRations)=R1min/R30sec
WhereR10min:Theinsulationresistancemeasuredatthe10minutepoint
afterpressingtheTESTbutton.
R1min:Theinsulationresistancemeasuredatthe1minutepointafter
pressingtheTESTbutton.
R30sec:Theinsulationresistancemeasuredatthe30secondpointafter
pressingtheTESTbutton.
OtherIndications ProgressiveLowBattery,OverRange