Product Manual

FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
24
Function ResistanceRange Resolution AccuracyofReading
Insulation
Resistance
4.000M
40.00M
0.001M
0.01M
±(1.5%+5digits)
400.0M
4000M
0.1M
1M
±(3.0%+5digits)
4.1GΩ‐20.0G 0.1G ±(10.0%+3digits)
Testvoltages Min.Resistance Max.Resistance
50V 50kΩ 50.0MΩ
100V 100kΩ 100.0MΩ
250V 250kΩ 250.0MΩ
500V 500kΩ 500.0MΩ
100V 1MΩ 20GΩ
Notes: 1.TestCurrent=1mA
2.TestVoltageAccuracy=0%,+20%ShortCircuitTestCurrent=1mA(nominal)
3.AutoDischargeFunction:Dischargetime<1secondforCap.<1µF
4.Max.CapacitiveLoad:Operablewithupto1µFload.
5.LiveCircuitDetection:If>2V
AC/DCdetectedtestinhibited
7.3InputSpecifications
Function MaximumInput
ACVoltage,DCVoltage 1000VDC/AC
InsulationTest NoLIVEVoltageIN‐Protected