User Manual
FLIR IM75 USER MANUAL Document Identifier: IM75-en-US_AB
21
6. Maintenance
6.1 Cleaning and Storage
Clean the meter with a damp cloth and mild detergent; do not use abrasives or solvents.
If the meter is not to be used for an extended period, remove the batteries and store them
separately.
6.2 Battery Replacement
The Battery symbol flashes with no ‘bars’ when the batteries have reached a critical level of 7.0V.
The symbol stays active and visible while the LCD is powered.
The meter displays readings within specifications while the low battery indicator is on. When this is
no longer possible, the display must be blanked. The meter powers off before it displays an out of
tolerance voltage.
1. To avoid electrical shock, disconnect the meter if connected to a circuit, remove the probe
from the terminals, and set the function switch to the OFF position before attempting to
replace the batteries.
2. Unscrew and remove the battery compartment cover.
3. Replace the six standard AAA batteries, observing correct polarity.
4. Secure the battery compartment cover.
6.3 Fuse Replacement
The fuse accessed via the battery compartment cover. The fuse is rated 440mA/1000V, ceramic
fast blow with a minimum Interrupt Rating of 10kA.
6.4 Disposal of Electronic Waste
As with most electronic products, this equipment must be disposed of in an
environmentally friendly way, and in accordance with existing regulations for
electronic waste. Please contact your FLIR Systems representative for more details.