Moisture Measurement Operation
7
Reference mode is selected, the current reading is offset to zero. Subse-
quent readings will be offset by this same amount, see Section 7.8 Set
Reference Mode for additional information.
Figure 7.13 Selecting the ‘Set Reference’ mode.
7.5 Pinless Moisture Measurements
Select the pinless mode from the menu system (Moisture mode/Pinless);
see Figure 7–11 above).
Place the internal moisture sensor (back) on the surface of the material to
be tested. Apply light pressure to ensure that the internal sensor is com-
pletely flat against the surface of the material under test.
The internal moisture sensor detects moisture to a depth of approximately
19 mm (0.75”). The actual depth will vary depending upon the amount of
moisture, the material under test, surface roughness, and other factors.
Pinless measurement readings are ‘relative’ scaled (0~100).
The moisture reading is displayed on the main display in the upper left
hand corner (thermal and camera image modes) or as large digits (Mois-
ture-only mode). Refer to Section 5.4 Image Modes and Sections 7.2
through 7.4, for more information.
Keep hands, surfaces, and objects away from the internal moisture sensor
on the back of the MR277 when taking measurements.
For best results, lift the meter off the surface under test between measure-
ment points; do not drag the meter over surfaces.
7.6 Pin Probe Measurements
1. Select the Pin mode from the menu (Moisture mode/Pin Mode); see Fig-
ure 7–10 above.