User Manual Ball Probe Moisture Meter With Bluetooth® Model MR59
Table of Contents 1. Advisories ................................................................................................................... 3 1.1 Copyright...................................................................................................................................... 3 1.2 Quality Assurance .......................................................................................................................... 3 1.3 Documentation ...............................................
1. Advisories 1.1 Copyright © 2018, FLIR Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual or otherwise, without the prior written permission of FLIR Systems.
2. Introduction Thank you for selecting the FLIR MR59 Ball Probe Moisture Meter with Bluetooth®. This instrument detects moisture in wood and other building materials and transmits readings to a remotely paired smart device. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. 2.
3.2 Display Description 1. Bluetooth® (BLE) icon (when paired) 2. Transmission icon (flashes when paired) 3. Data Hold icon 4. Alarm enabled icon 5. Battery status icon 6. Maximum reading 7. Minimum reading 8. Moisture reading 3.
4. Operation 4.1 Meter Power, APO, and Zero Calibration 1. One (1) 9V battery (rear compartment) powers the meter. 2. Long press the power button to switch the meter ON/OFF. If the meter display does not switch ON, please check the battery. 3. The meter beeps and flashes CAL while it performs a self-test (zero calibration). Keep hands and objects away from the sensor while the meter powers up for optimum calibration accuracy. 4.
4.7 High Moisture Audible Alarm The MR59 beeps when the moisture reading exceeds the high alarm limit. To set the alarm limit, short press the SET button and use the arrow buttons to set the value. To exit the alarm programming screen, short press the SET button again. To enable/disable the audible alarm, long press the SET button. When the alarm is enabled (default), the alarm bell icon appears. The alarm limit is set to ‘20’ by default. 4.
6. Field Firmware Upgrades via USB Interface The MR59 includes a micro USB port, located inside the battery compartment. The USB port allows the user to upgrade the System firmware or the Bluetooth® firmware by first downloading an upgrade file from the FLIR website and then connecting the meter to a PC to transfer the file to the meter. Firmware upgrades are available at the http://support.flir.com website.
13. When the upgrade is complete, the meter will return to the normal operating mode. 6.2 Bluetooth® Firmware Upgrade To execute a Bluetooth® Firmware Upgrade please perform the same steps as listed in Section 6.1, System Firmware Upgrade, above. If there is an error, the ‘Err’ icon will flash and the meter will return to the normal operation mode. If this error occurs, please repeat the upgrade procedure. If the problem persists, contact FLIR technical support. 7. Maintenance 7.
8. Safety Handle the meter carefully and do not subject the sensor to shock or excessive vibration. Keep the instrument dry. Pack the battery separately when the meter is to be stored for periods longer than 60 days. Do not use this device for purposes other than those described in this manual. 8.1 FCC Compliance This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: 1. This device may not cause harmful interference. 2.
Low battery indicator Operating RH/Temperature Housing material Drop-proof IP rating Safety Compliance Dimensions Weight (100% to empty) 90%, 32~86°F (0~30°C) 75%, 86~104°F (30~40°C) 45%, 104~122°F (40~50°C) Impact resistant plastic 6.6 ft. (2m) IP40 CE, RCM 9.5 x 2.6 x 1.5 in. (240.5 x 67 x 38mm) 86 oz. (245g) 10. Technical Support Main Website http://www.flir.com/test Technical Support Website http://support.flir.com Technical support Email TMSupport@flir.
Corporate Headquarters FLIR Systems, Inc. 2770 SW Parkway Avenue Wilsonville, OR 97070 USA Telephone: +1 503-498-3547 Customer Support Technical Support Website Technical Support Email Service and Repair Email Customer Support Telephone http://support.flir.com TMSupport@flir.com Repair@flir.com +1 855-499-3662 option 3 (toll free) Publication Identification No.