User Manual

FLIR MR60 USER MANUAL Document Identifier: MR60-en-US_AB
13
5.4 High Moisture Alarm
The MR60 offers a High Moisture Alarm where an audible and visual alert activates when the
moisture reading exceeds the programmed high limit.
1. Press the Select button to access the Main Menu
2. Select the SETTINGS mode from the Main Menu
3. Scroll to ALARM and press the Select button to open the Alarm programmer
4. Use the Navigation and Select buttons to switch the alarm ON or OFF and to set the
threshold from 0% to 100%
5. Press the Select button to return to the SETTINGS mode and save the value or press the
back button to cancel and to return to the normal operating mode
6. When the High Alarm is set ON, the main display will show the alarm bell icon and the
High Alarm value. Refer to the example screen in Fig. 5-1.
7. When the measurement exceeds the threshold, the bargraph will turn red in color and
the meter will beep.
8. To switch the alarm OFF when the meter is alarming, press the Select button twice. The
meter will then be in the Settings mode where the alarm can be reprogrammed if
desired.
Fig. 5-3 Log file example