Operation Manual
Handling the camera13
5. Activate the global parameter External IR window compensation by doing the
following:
5.1. Push the joystick to display the menu system.
5.2. Select
(Measurement parameters) and push the joystick. This displays a
submenu.
5.3. Select
(External IR window compensation) and push the joystick. This dis-
plays a dialog box.
5.4. In the dialog box, define the settings for the infrared window compensation:
• Turn compensation on/off: Select On.
• Temperature: Select the temperature of the close-up lens.
• Transmission: Select 0.89, which is a typical transmission value for the
close-up lens.
5.5. Push the Back button
to go back in the menu system.
5.6. The status icon
is now displayed.
6. Select the image mode Thermal by doing the following:
6.1. Push the joystick to display the menu system.
6.2. Select
(Image mode) and push the joystick. This displays a submenu.
6.3. Select
(Thermal) and push the joystick.
6.4. Push the Back button
to go back in the menu system.
7. Deactivate continuous autofocus by selecting
(Settings) > Device settings > Con-
tinuous autofocus = Off.
13.14.3 Removing the close-up lens
Note Do not touch the lens surfaces when you remove the close-up lens. If this happens,
clean the lens according to the instructions in 31.2 Infrared lens, page 216.
Follow this procedure:
1. Turn the close-up lens counter-clockwise until it separates from the infrared lens.
#T559954; r. AL/37426/37426; en-US
44