Appendix C EMC Information
9.1 Electromagnetic Emissions
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 Class A). If used in a residential environment (for which CISPR 11 Class B is normally
required) this equipment may not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
Emissions Class and Group Compliance
The FloPatch FP120 is intended for use in the electromagnetic environment specified
below. The customer or the user of the FloPatch FP120 should assure that it is used in such
an environment.
Emissions Test
Compliance
Electromagnetic Emissions - Guidance
RF Emissions
CISPR 11
Group 1
The FloPatch FP120 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The FloPatch FP120 is suitable for use in all establishments
other than domestic, and may be used in domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following
warning is heeded:
Warning: This equipment/system is intended for use by
trained medical professionals only. This equipment/system
may cause radio interference or may disrupt the operation
of nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or relocating the
FloPatch FP120 or shielding the location.
Harmonic Emissions
IEC 610000-3-2
Not Applicable
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not Applicable
9.2 Electromagnetic Immunity
The FloPatch FP120 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FloPatch FP120 should assure that it is used in such an
environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD) IEC
61000-4-2
±8 kV Contact
±15 kV Air
±8 kV Contact
±15 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC
61000-4-4
±2 kV for power
supply lines
Not Applicable
Not Applicable