13.11.4 Direct simulation of outputs
If the operator selected the setting “Direct simulation” described in Section 13.11.2 Direct simulation on
page 86, the following four possible settings will affect the output behavior during measured value simula-
tion, where all measured values are simulated at the same time.
13.11.4.1 Status output simulation
The
Status output simulation function allows the operator to purposefully activate the status output.
Press
to display the current state.
Status output
[off]
As mentioned in Section 12.4.3.1 Selection window/make a selection, the operator can toggle between
“on” and “off.”
13.11.4.2 Pulse output simulation
The
Pulse output simulation function allows the operator to define a frequency to be assigned to the pulse
output. After selecting this function and pressing
, the following selection field will be displayed:
Set frequency
0210.0 Hz
This field shows the current frequency. As mentioned in Section 12.4.3.2 Input window/modify a value,
the definable frequency ranges from 6 Hz to 1100 Hz.
13.11.4.3 Simulation current output I1
This function allows the operator to define a current for current interface 1. Press
to display the set
current.
Set I1
I1 = 10.50 mA
As mentioned in Section 12.4.3.2 Input window/modify a value, the current value can be changed.
13.11.4.4 Simulation current output I2
As described in Section 13.11.4.3, current output 2 can also be configurated.