Electromagnetic Compatibility
en
11.5 Guidelines and Manufacturer's Statement/Electromagnetic In-
terference Immunity
The device FLUID CONTROL Lap 2216 is intended for operation in an electromag-
netic environment as described below. The user of the device FLUID CONTROL Lap
2216 should make sure the device is operated within such an environment.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
Electromagnetic
interference
immunity tests
IEC 60601-1-
2 test level
Compliance
levels
Electromagnetic environ-
ment/guidelines
Conducted HF
interference
quantities
according to
IEC 61000-4-6
Radiated HF
interference
quantities
according to
IEC 61000-4-3
3 V
eff
150 kHz to
80 MHz
3 V/m 80
MHz to
2.7 GHz
3 V
eff
80 MHz to
800 MHz
3 V/m
80 MHz to
2.7 GHz
Portable and mobile wire-
less devices should not be
used in closer proximity to
the device FLUID CONTROL
Lap 2216 (including
cables/lines) than the rec-
ommended safety distance
calculated based on the
transmitting frequency
and the applicable for-
mula. Recommended
safety distances:
d=1.2P for 150 KHz
to 80 MHz
d=1.2P for 80 MHz to
800 MHz
d=2.3P for 800 MHz
to 2.7 GHz
P is the max. rated output
of the transmitter in watt
(W) according to the infor-
mation provided by the
manufacturer of the trans-
mitter.
The recommended separa-
tion distance in meters (m)
is d.
The field strength of sta-
tionary transmitters for all
frequencies tested on site
a
should be lower than the
concordance level.
b
Interference is possible in
the proximity of devices
featuring the following
pictograph.