Infrared Thermometer
Temperature Measurement
Range ................................................. -20 °C to 200 °C (-4 °F to 392 °F)
Accuracy (typical)
-20 °C to 120 °C
(-4 °F to 248 °F) .............................. ±2 °C (4 °F)
120 °C to 160 °C
(248 °F to 320 °F) ........................... ±3 °C (6 °F)
160 °C to 200 °C
(320 °F to 392 °F) ........................... ±4 °C (7 °F)
Note
The specified accuracy is applicable only when the Meter is in
thermal equilibrium with its environment. Accuracy is not
specified when the target temperature is more than 20
°
C
(36
°
F) below the temperature of the Meter.
Focal length ........................................ Fixed, at ~3.8 cm (1.5 in)
External Sensor
Frequency Range ............................... 10 Hz to 1,000 Hz
Bias Voltage (to supply power) ........... 20 V DC to 22 V DC
Bias Current (to supply power) ........... 5 mA
Note
Fluke supports, but does not provide, external sensors.
Vibration Meter
Low Frequency Range
(overall measurement) ........................ 10 Hz to 1,000 Hz
High Frequency Range
(CF+ measurement)............................ 4,000 Hz to 20,000 Hz
Vibration Limit .................................... 50 g peak (100 g peak-peak)
Battery Type ....................................... AA (2) lithium non-rechargeable
3 V dc
Battery Life
805 .................................................. 250 measurements
805 FC, Fluke Connect enabled .. 210 measurements
A/D Converter ..................................... 16-bit
Sampling Rate
Low Frequency ............................... 20,000 Hz
High Frequency ............................... 80,000 Hz
Signal to Noise Ratio .......................... 80 dB
Real Time Clock Backup ..................... Coin Battery
Size (L x W x H) .................................. 24.1 cm x 7.1 cm x 5.8 cm (9.5 in x
2.8 in x 2.3 in)
Weight ................................................. 0.40 kg (0.89 lb)
Connectors (805) ................................ USB Mini-B 7-pin, Stereo Audio
Output Jack (3.5 mm Audio Plug),
External Sensor Jack (SMB
connector)
Connectors (805 FC) .......................... USB Mini-B 7-pin, External Sensor
Jack (SMB connector)
Firmware
External Interfaces .............................. USB 2.0 (full speed)
communication
Data Capacity ..................................... Database on internal flash memory
Upgrade .............................................. through USB
Memory ............................................... Up to 3,500 measurements
Environmental
Operating Temperature ....................... -20 °C to 50 °C (-4 °F to 122 °F)
Storage Temperature .......................... -20 °C to 60 °C (-4 °F to 140 °F)
Operating Humidity ............................. 10 % to 95 % RH (non-