12
Device Info
Information about your Meter is in the Device Info menu.
This information includes the:
Serial number
Software version
Emissivity value (See Emissivity Selection for more
information on the emissivity value.)
Internal sensor sensitivity
Database Free Space
Battery Selection
To change the battery type:
1. Go to the Device Settings menu.
2. Push and to highlight Battery Selection.
3. Push to open the menu.
4. Push and to highlight the battery type you
have in the Meter.
5. Push to update the Meter and exit the
menu.
Emissivity Selection
The correct emissivity value is important for you to make
the most accurate temperature measurements. Most
painted or oxidized surfaces have an emissivity of 0.93
(default value set in the Meter). This is correct for non-
contact temperature measurements on most bearing
housings.
Inaccurate measurements can result from shiny or
polished metal surfaces. To compensate, put masking
tape or flat black paint on the measurement surface. Make
sure that the tape is the same temperature as the
measurement surface before you make a measurement.
For other applications the Meter has additional preset
emissivity values:
Aluminum (e=0.30)
Iron (e=0.70)
Steel (e=0.80)
Paint (e=0.93) – default value
Wood (e=0.94)