34
Fuse Replacement
To replace the fuse:
1. Remove the test leads from the Meter.
2. Remove the display module from the Meter base.
See the “Remove the Display Module” section.
3. As shown in Figure 12, remove four screws from the
case bottom.
4. Pull the case bottom from the case top.
5. Remove the fuse from its holder and replace it with
an 11 A, 1000 V, FAST fuse with a minimum interrupt
rating of 17,000 A. Use only Fluke PN 803293.
To re-assemble the Meter, do the steps above in the
opposite sequence.
3
5
4
gcc113.eps
Figure 12. Fuse Replacement