User's Manual

114, 115, 116, and 117 Calibration Information Manual Supplement
11/09 1
Change #1
On page 10, delete step 30.
Change #2
On page 4, replace Table 2 with the following:
Table 2. Accuracy Specifications
Function Range Resolution
Accuracy
([% of Reading] + [Counts])
Model
[1]
DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2 114, 115, 116, 117
DC, 45 to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ
[2]
True-rms 600.0 V 0.1 V 2.0 % + 3 4.0 % + 3 114, 116, 117
45 to 500 Hz 500 Hz to 1 kHz
AC millivolts
[2]
True-rms 600.0 mV 0.1 mV 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
AC Volts
[2]
True-rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Continuity 600 1
Beeper on <20 , off >250 ; detects opens
or shorts of 500 s or longer.
114, 115, 116, 117
Ohms
600.0
6.000 k
60.00 k
600.0 k
6.000 M
40.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
5.0 % + 2
114, 115, 116, 117
Diode test 2.000 V 0.001 V 0.9 % + 2 115, 116, 117
Capacitance 1000 nF
10.00 F
100.0 F
9999 F
1 nF
0.01 F
0.1 F
1 F
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 F - 1000 F: 1.9 % + 2
>1000 F: 5 % + 20
115, 116, 117
Lo-Z Capacitance
(Power-up option)
1 nF to 500 F 10 % +2 typical
115, 116, 117
AC Amps True-rms
[2]
(45 Hz to 500 Hz)
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.5 % + 3 115, 117
DC Amps
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.0 % + 3 115, 117