User's Manual

Manual Supplement 114, 115, 116, and 117 Calibration Information
2 4/13
Temperature (Type K
thermocouple)
-40 C to 400 C
-40 F to 752 F
0.1 C
0.2 F
1.0 % + 10
[5]
1.0 % + 18
[5]
116
AC Amps True-rms
[2]
(45 Hz to 1 kHz)
600.0 A 0.1 A
1.5 % + 3 (2.5 % + 3 > 500 Hz) 116
DC Amps 600.0 A 0.1 A 1.0 % + 2 116
Hz (V or A input)
[3]
99.99 Hz
999.9 Hz
9.999 kHz
50.00 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 2 115, 116, 117
Notes:
[1] Models listed in this column also refer to the “C” version of the model. For example, those rows containing model 115 are applicable to the 115C
as well.
[2] All ac ranges except Auto-V LoZ are specified from 1 % to 100% of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range
are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit
or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is
ac-coupled. Auto-V LoZ, AC mV, AC amps, and AC amps are dc-coupled.
[3] AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC amps Hz is dc-coupled and specified from 45 Hz to 5 kHz.
[4] >10 A unspecified.
[5] Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
Change #3
On page 2, under Safety Information:
Change:
Do not use the Meter around explosive gas or vapor.
To: Do not use the Product around explosive gas, vapor, or in damp or wet environments
Change #4, 63151, 63405
On page 3, replace Table 1 with:
Table 1. Electrical Symbols
Symbol Description Symbol Description
AC (Alternating Current) Fuse
DC (Direct Current) Double Insulated
Hazardous voltage Important Information; Refer to manual
Battery (Low battery when shown on the
display.)
Earth ground
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
Conforms to relevant South Korean EMC Standards
CAT II
Measurement Category II is applicable to test and measuring circuits connected directly to utilization points
(socket outlets and similar points) of the low-voltage MAINS installation.
CAT III
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of
the building’s low-voltage MAINS installation.
CAT IV
Measurement Category IV is applicable to test and measuring circuits connected at the source of the
building’s low-voltage MAINS installation.