2. Press and select MEASURE on A: CAP .
Verify TP156 for +3.3 ... 0V pulses (repetition rate 100...200 ms):
Zero scale (open input): pulse width approximately 30 µs.
Full scale (for example 500 nF): pulse width approximately 25 ms.
If not correct, most probably the C-ASIC N101 is defective.
If correct, continue at Section 7.5.8 Trigger functions (pulse width is measured via
the T-ASIC).
7.5.8 Trigger Functions
1. Press and select MEASURE on A: VDC .
2. Press and select INPUT A: DC | NORMAL | INPUT B: DC | NORMAL
3. Press to select the SCOPE INPUTS menu.
Press
to select the TRIGGER menu, and select:
INPUT: A or B | SCREEN UPDATE: FREE RUN | AUTO RANGE: . >15HZ
Press
to open the SCOPE INPUTS menu.
Press
to open the SCOPE OPTIONS ... menu, and select:
SCOPE MODE: NORMAL | WAVEFORM MODE: NORMAL.
4. Supply a 1 kHz sine wave of +/- 3 divisions to Input A, and Input B.
5. Check:
a. TP156, TP256 for a 600 mV (6 div. x 100 mV/div), 1 kHz, sine wave; the DC
level depends on the trace position.
If not correct, C-ASIC N101/N102 is probably defective.
b. TP321, TP322 for 1.1...1.9V DC (move the trigger level from top to bottom).
If not correct check the PWM circuit, see Section 7.5.8.
c. TP311for a 0...+3.3V, 1 kHz square wave when the trigger level is at the middle
of the trace). Change the trigger level, and verify that the duty cycle of the
square wave changes. If not correct T-ASIC N301 may be defective.
d. TP433 for 0...+3.3V pulses. Pulse width:
4...10 µs for time base 2 µs/div and faster;
>40 µs for time base 5 µs/div and slower; pulse width increases with time base.
e. TP336 for +0.6...0V pulses, TP436 for +3.3...0V pulses; the pulse width is about
40 µs...10 ms.
If not correct, check the RANDOMIZE circuit, see Section 7.5.15.
f. TP437 (SMPCLK) for a 5 MHz (time base 10 ms/div) or 25 MHz (time base <
10 ms/div) clock signal (3.3V). Check SMPCLK on both sides of R339.
6. To test video trigger press
to select the SCOPE INPUTS menu.
Press to select the TRIGGER menu, and select INPUT: VIDEO on A...
From the VIDEO TRIGGER submenu select:
SYSTEM: PAL | LINE: RANDOM | POLARITY: POSITIVE
Press
to open the SCOPE INPUTS menu.
Press
to open the SCOPE OPTIONS ... menu, and select:
SCOPE MODE: NORMAL | WAVEFORM MODE: NORMAL