User's Manual

123
Service Manual
3-18
Table 3-3. Voltage Ranges And Trace Sensitivity
range 50 mV 50 mV 50 mV 500 mV 500 mV 500 mV 5V 5V
trace ../div 5 mV 10 mV 20 mV 50 mV 100 mV 200 mV 500 mV 1V
range 5V 50V 50V 50V 500V 500V 500V 1250V
trace ../div 2V 5V 10V 20V 50V 100V 200V 500V
During measuring, input voltage measurements, gain measurements, and zero
measurements are done. As a result, the voltage supplied to the ADC is a multiplexed
(zero, + reference, -reference, input voltage) signal. In ROLL mode however, no gain
and zero measurements are done. Now the ADC input voltage includes only the
conditioned input voltage.
The input voltage is connected to Input A. The shield of the input is connected to system
ground (
) via a PTC ground protection resistor. If a voltage is applied between the
Input A and Input B ground shield, or between one of these ground shields and the black
COM input, the PTC resistor will limit the resulting current. The voltage across the PTC
resistor is supplied to the C-ASIC GPROT input, and causes a ground error warning
(high voltage level) on output pin 24 (DACTEST).
Resistance Measurements (Channel A)
The unknown resistance Rx is connected to Input A, and the black COM input. The T-
ASIC supplies a constant current to Rx via relay contacts K173, and the PTC resistor
R172. The voltage across Rx is supplied to a high impedance input buffer in the C-ASIC
via the LF input pin 42. The C-ASIC conditions the voltage across Rx to an output
voltage of 50 mV/div. This voltage is supplied to the ADC on the Digital part. The
ADC data is read and processed by the D-ASIC, and represented as a numerical reading,
and a graphical trace in a fixed time base.
Table 3-4 shows the relation between the reading range (), the trace sensitivity
(/div.), and the current in Rx. The selected trace sensitivity determines the C-ASIC
attenuation/gain factor. The reading range is only a readout function, it does not change
the hardware range or the wave form display.
Table 3-4. Ohms Ranges, Trace Sensitivity, and Current
Range 50 500 5k 50 k 500 k 5 M 30 M
Sensitivity ../div 20 200 2 k 20 k 200 k 2 M 10 M
Current in Rx 500 µA 500 µA 50 µA5 µA 500 nA 50 nA 50 nA
To protect the current source from being damaged by a voltage applied to the input, a
PTC resistor R172 and a protection circuit are provided (See Section 3.3.3 “Current
Source”).
During measuring, input voltage measurements, gain measurements, and zero
measurements are done. As a result, the voltage supplied to the ADC is a multiplexed
(zero, + reference, -reference, input voltage) signal.
Capacitance Measurements (Channel A)
The capacitance measurement is based on the equation: C x dV = I x dt. The unknown
capacitor Cx is charged with a constant known current. The voltage across Cx increases,