User's Manual

Fluke 123/124
Getting Started
2
Warning
To avoid electrical shock or fire:
Use only the power supply, Model PM8907
(Battery Charger / Power Adapter).
Before use check that the selected/indicated
voltage range on the PM8907 matches the
local line power voltage and frequency.
For the PM8907/808 universal Battery
Charger/Power Adapter use only line cords
that comply with the local safety regulations.
Note
To accommodate connection to various line power
sockets, the PM8907/808 universal Battery Charger /
Power Adapter is equipped with a male plug that must
be connected to a line cord appropriate for local use.
Since the adapter is isolated, the line cord does not
need to be equipped with a terminal for connection to
protective ground. Since line cords with a protective
grounding terminal are more commonly available you
might consider using these anyhow.
The 230V rating of the PM8907/808 is not for use in
North America. A line plug adapter complying with the
applicable National Requirements may be provided to
alter the blade configurations for a specific country.
Warning
To avoid electrical shock or fire if a Test Tool
input is connected to more than 42 V peak (30
Vrms) or on circuits of more than 4800 VA:
Use only insulated voltage probes, test leads
and adapters supplied with the Test Tool, or
indicated as suitable for the Fluke 123/124
Test Tool.
Before use, inspect voltage probes, test leads
and accessories for mechanical damage and
replace when damaged.
Remove all probes, test leads and
accessories that are not in use.
Always connect the battery charger first to
the ac outlet before connecting it to the Test
Tool.
Do not apply input voltages above the rating
of the instrument. Use caution when using 1:1
test leads because the probe tip voltage will
be directly transmitted to the Test Tool.
Do not use exposed metal BNC or banana
plug connectors.
Do not insert metal objects into connectors.
Always use the Test Tool only in the manner
specified.