User's Manual
Fluke 123/124
Getting Started
2
Warning
To avoid electrical shock or fire:
• Use only the power supply, Model PM8907
(Battery Charger / Power Adapter).
• Before use check that the selected/indicated
voltage range on the PM8907 matches the
local line power voltage and frequency.
• For the PM8907/808 universal Battery
Charger/Power Adapter use only line cords
that comply with the local safety regulations.
Note
To accommodate connection to various line power
sockets, the PM8907/808 universal Battery Charger /
Power Adapter is equipped with a male plug that must
be connected to a line cord appropriate for local use.
Since the adapter is isolated, the line cord does not
need to be equipped with a terminal for connection to
protective ground. Since line cords with a protective
grounding terminal are more commonly available you
might consider using these anyhow.
The 230V rating of the PM8907/808 is not for use in
North America. A line plug adapter complying with the
applicable National Requirements may be provided to
alter the blade configurations for a specific country.
Warning
To avoid electrical shock or fire if a Test Tool
input is connected to more than 42 V peak (30
Vrms) or on circuits of more than 4800 VA:
• Use only insulated voltage probes, test leads
and adapters supplied with the Test Tool, or
indicated as suitable for the Fluke 123/124
Test Tool.
• Before use, inspect voltage probes, test leads
and accessories for mechanical damage and
replace when damaged.
• Remove all probes, test leads and
accessories that are not in use.
• Always connect the battery charger first to
the ac outlet before connecting it to the Test
Tool.
• Do not apply input voltages above the rating
of the instrument. Use caution when using 1:1
test leads because the probe tip voltage will
be directly transmitted to the Test Tool.
• Do not use exposed metal BNC or banana
plug connectors.
• Do not insert metal objects into connectors.
• Always use the Test Tool only in the manner
specified.