550C/1555 Insulation Tester Calibration Manual September 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Limited Warranty and Limitation of Liability Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Introduction........................................................................................................ How to Contact Fluke ........................................................................................ Safety Information ............................................................................................. Symbols ............................................................................................................. Required Equipment ..........
1550C/1555 Calibration Manual ii
List of Tables Table 1. 2. 3. 4. 5. 6. 7. Title Symbols.................................................................................................................. Required Equipment............................................................................................... Insulation Accuracy Test........................................................................................ Output Voltage Test .............................................................................................
1550C/1555 Calibration Manual iv
List of Figures Figure Title 1. 2. 3. 4. Pushbuttons ............................................................................................................ Display Features..................................................................................................... IR Port .................................................................................................................... Query Constant Result .................................................................................
1550C/1555 Calibration Manual vi
Introduction Warning To prevent possible electrical shock, fire, or personal injury: • Do not perform the verification tests or calibration procedures described in this manual unless qualified to do so. • Read all safety Information before you use or service the product. The Calibration Manual for the 1550C/1555Insulation Tester (hereafter referred to as "the product") provides the following information.
1550C/1555 Calibration Manual How to Contact Fluke To contact Fluke, call one of the following telephone numbers: • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) • Canada: 1-800-36-FLUKE (1-800-363-5853) • Europe: +31 402-675-200 • Japan: +81-3-3434-0181 • Singapore: +65-738-5655 • Anywhere in the world: +1-425-446-5500 Or, visit Fluke's website at www.fluke.com. To register your product, visit http://register.fluke.com.
Insulation Tester Safety Information • Do not use test leads if they are damaged. Examine the test leads for damaged insulation, exposed metal, or if the wear indicator shows. Check test lead continuity. • Do not use and disable the product if it is damaged. • Do not touch voltages >30 V ac rms, 42 V ac peak, or 60 V dc. • Keep fingers behind the finger guards on the probes.
1550C/1555 Calibration Manual Symbols Symbols on the product and in the manual are explained Table 1. Table 1. Symbols Symbol Conforms to European Union standards. Canadian Standards Association is the certified body used for testing compliance to safety standards. Risk of Danger. Important information. See Manual. Hazardous voltage Equipment protected by double or reinforced insulation. Examined and licensed by TÜV Product Services.
Insulation Tester Required Equipment Required Equipment Equipment required to perform the procedures in this manual is listed in Table 2. If the recommended models are not available, equipment with equivalent specifications may be substituted. Warning For safe operation and maintenance of the product, have an approved technician repair the product. Caution Do not attempt to use the 5500A, 5520A, or other standard calibrator for insulation and continuity resistance tests. Calibrator damage will result.
1550C/1555 Calibration Manual Equipment Personal computer Minimum Required Characteristics Recommended Model IBM compatible, with Microsoft Windows XP SP2 or later + .NetFramework 2.0 or later 1550C/1555Users Manual CD ROM Fluke P/N 3592810 1. Resistors must have a voltage coefficient consistent with the test voltage used. 2. Can use (3) each, 0.033 μF, 2 kV capacitors in series and (8) each, 8 μF, 450 V capacitors in series to obtain required values. The 0.
Insulation Tester Performance Test Procedures Button Test Use the pushbuttons to control the product, view test results, and scroll through chosen test results. Pushbuttons and their functionality are discussed in Figure 1. FUNCTION 1 2 3 4 5 ENTER TEST 6 7 ghh02.eps Item Description Turns the product off and on. Push to go to the Function menu. Push again to exit the Function menu. To scroll within the Function menu, use the arrow pushbuttons.
1550C/1555 Calibration Manual Display Test Turn the UUT on several times while observing the display during power up. Compare the display with the example in Figure 2. Check all segments for clarity and contrast. 6 5 7 4 8 3 9 2 1 10 11 12 gHH03f.EPS Item Description Interference present. Readings could be out of specified accuracy range.
Insulation Tester Performance Test Procedures Charging Test 1. With the product switched off, connect a mains supply to the ac supply receptacle and check that the UUT display shows Charging. 2. Disconnect the mains supply and check that the UUT turns off. 3. Turn the product on and see that all the battery symbol segments display as shown in Figure 2. Note A fully-charged battery is indicated when the battery symbol shows all segments. Recharge the battery as necessary to obtain all segments.
1550C/1555 Calibration Manual The capacitance reading is obtained by pressing after a test has started. Table 3. Insulation Accuracy Test Step Voltage Range Minimum Maximum 1 250 V 0.1 μF 0.055 0.145 2 500 V 250 kΩ 237 kΩ 263 kΩ 3 500 V 1 GΩ 0.95 GΩ 1.05 GΩ 4 500 V 100 GΩ 80 GΩ 120 GΩ 5 1 kV 1 GΩ 0.95 GΩ 1.05 GΩ *6 2.1 kV 1 μF 0.82 1.18 7 2.5 kV 1 GΩ 0.95 GΩ 1.05 GΩ 8 5 kV 1 GΩ 0.95 GΩ 1.05 GΩ 9 5 kV 100 GΩ 95 GΩ 105 GΩ 10 5 kV 1 TΩ 0.80 TΩ 1.
Insulation Tester Performance Test Procedures Output Voltage Test In Table 4, the UUT output voltage is checked with various loads applied. In this test a voltmeter with a high-voltage probe must be connected to the load resistor to measure the UUT output voltage. Use 15 80K-6 for voltages below 6 kV. Table 4.
1550C/1555 Calibration Manual Voltage Measurement Accuracy To verify voltage measurement accuracy of the Live Circuit Warning function, apply the voltages listed in Table 6 to the + and - terminals of the UUT. Verify that: • The UUT reading is within the display limits of Table 6. • The UUT is beeping at a 1-second interval. • is flashing on the display. Table 6.
Insulation Tester Adjustment Procedure Instrument Setup Turn the product on and wait for Test Voltage to appear on the display. From the computer terminal, activate the Snorre program from the Windows Start menu. On the Setup tab, confirm the selected COM port settings. Warning To prevent possible electrical shock, fire, or personal injury, do not contact the output terminals or test equipment terminals while performing the following procedures.
1550C/1555 Calibration Manual Current Adjustment 1. Select the Cal Current tab. 2. Attach a 2 mA current source to the LO and GUARD terminals of the UUT, connecting the current source LO to UUT GUARD terminal, as shown in the connection diagram. 3. Apply 2 mAdc to the UUT. 4. Press the START button and wait until the adjustment is complete. 5. The current measurement is now adjusted. Disconnect the current source. Charge Adjustment 1. Select the Cal Charge tab. 2.
Insulation Tester Additional Procedures Get Diagnostics Pressing this button continually gets Raw ADC values from the UUT and updates the Raw ADC Counts boxes (v_counts, i_counts, q_counts), pressing the button again turns this feature off. Query Constants Pressing this button provides an html screen dump of the present opvars obtained from the UUT. You will need to check that numerical values are reported for all nine variables. See Figure 4. Figure 4. Query Constant Result snorre 3.
static awareness A Message From Fluke Corporation Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by: 1. Knowing that there is a problem. 2. Learning the guidelines for handling them. 3. Using the procedures, packaging, and bench techniques that are recommended. The following practices should be followed to minimize damage to S.S. (static sensitive) devices. 1.
5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT. 8. WHEN REMOVING PLUG-IN ASSEMBLIES HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES. 6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE. 9. HANDLE S.S. DEVICES ONLY AT A STATIC-FREE WORK STATION. 10. ONLY ANTI-STATIC TYPE SOLDERSUCKERS SHOULD BE USED. 11. ONLY GROUNDED-TIP SOLDERING IRONS SHOULD BE USED. 7.
1550C/1555 Calibration Manual Battery Replacement Procedure Warning For safe operation and maintenance of the product: • Batteries contain hazardous chemicals that can cause burns or explode. If exposure to chemicals occurs, clean with water and get medical aid. • Remove all probes, test leads, and accessories before the battery door is opened. • Remove all probes, test leads, and accessories before the case is opened. • The battery door must be closed and locked before you operate the product.
Insulation Tester Battery Replacement Procedure Note Pb This instrument contains a lead-acid battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler. Contact your authorized Fluke Service Center for disposal and recycling information. Storing rechargeable lead-acid batteries in a low-charged state could lead to reduced life and/or damage.
1550C/1555 Calibration Manual Cleaning Warning For safe operation and maintenance of the product, remove excess water from the cloth before cleaning the product to ensure that water does not enter any terminal. Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents to clean the product. Replacement Parts/Accessories Replacement parts are listed in Table 7. Table 7.
Insulation Tester Specifications Specifications General Specifications Display Power 75 mm x 105 mm 12 V lead-acid rechargeable battery. 2.6 Ahr 85 V to 250 V ac, 50/60 Hz, 20 VA Charger Input (AC) Dimensions (H x W x L) This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only. 170 mm x 242 mm x 330 mm (6.7 in. x 9.5 in. x 13.0 in.) Weight 3.
1550C/1555 Calibration Manual Electrical Specifications Product accuracy is specified for 1 year after calibration at operating temperatures of 0 °C to 35 °C. For operating temperatures outside the range (-20 °C to 0 °C and 35 °C to 50 °C), add ±0.25 % per °C, except on the 20 % bands add ±1 % per °C.
Insulation Tester Principle of Measurement and Resistance Range Leakage Current Measurement Capacitive Measurement Accuracy 1 nA to 2 mA ±(20 % + 2 nA) 0.01 μF to 20.00 μF ±(15 % of reading + 0.
1550C/1555 Calibration Manual 22