User's Manual

Insulation Testers
Performance Tests
21
5. Press T, and verify that the UUT reading is within the display reading limits
shown in Table 9.
6. Connect the Fluke 5320A high resistance output terminals to the UUT
COM and
Ω terminals.
7. Set the Fluke 5320A to the high resistance source function (H).
8. Press T, and verify that the reading is within the display reading limits for
Table 9, Step 4.
Table 9. Earth Bond Resistance Tests
Display Reading
Step Tester Range Applied Display Units
Low Limit High Limit
1 20.00 Ω 2.0 Ω Ω 1.94 2.06
2 2000 Ω 810.0 Ω Ω 795 825
3 2000 Ω 990.0 Ω Ω 972 1008
4 20.00 kΩ 18.0 kΩ kΩ 17.70 18.30
2-Ohm Output Current Test
The following procedure verifies the minimum current for a 2-Ω Load in continuity
function:
1. Connect the Fluke 5320A low resistance output terminal to the UUT
COM and Ω
terminals.
2. Set the Fluke 5320A to the low resistance source function (L) and 2-Ω output.
3. Set the UUT rotary function switch to the Ω function.
4. Press the UUT T. The Fluke 5320A current reading should be > 200 mA.
Open Circuit Voltage Test
The following test confirms that the open circuit voltage for the UUT Ω function is
within limits.
1. Connect the Fluke 5320A
V and COM meter terminals directly to the UUT Ω and
COM terminals observing correct polarity.
2. Set the Fluke 5320A to the M mode and dc volts.
3. Set the UUT rotary function switch to the
Ω function.
4. Press T and verify that the Fluke 5320A voltage reading is > 4.0 V but
< 8.0 V.