Data Sheet

5 Fluke Corporation 1587/1577 Insulation Multimeters
Model 1587
Output Voltage Display Range Resolution Test Current Resistance Accuracy
50 V
(0 % to + 20 %)
0.01 to 6.00 M 0.01 M
1 mA @ 50 k ± (3 % + 5 counts)
6.0 to 50.0 M 0.1 M
100 V
(0 % to + 20 %)
0.01 to 6.00 M 0.01 M
1 mA @ 100 k ± (3 % + 5 counts)6.0 to 60.0 M 0.1 M
60 to 100 M 1 MΩ
250 V
(0 % to + 20 %)
0.1 to 60.0 M 0.1 M
1 mA @ 250 k ± (1.5 % + 5 counts)
60 to 250 M 1 MΩ
500 V
(0 % to + 20 %)
0.1 to 60.0 M 0.1 M
1 mA @ 500 k ± (1.5 % + 5 counts)
60 to 500 M 1 MΩ
1000 V
(0 % to + 20 %)
0.1 to 60.0 M 0.1 M
1 mA @ 1 M
± (1.5 % + 5 counts)
60 to 600 M 1 MΩ
0.6 to 2.0 G 100 M ± (10 % + 3 counts)
Model 1577
Output Voltage Display Range Resolution Test Current Resistance Accuracy
500 V
(0 % to + 20 %)
0.1 to 60.0 M 0.1 M
1 mA @ 500 k ± (2.0 % + 5 counts)
60 to 500 M 1 MΩ
1000 V
(0 % to + 20 %)
0.1 to 60.0 M 0.1 M
1 mA @ 1 M ± (2.0 % + 5 counts)
60 to 600 M 1 MΩ
General specifications
Maximum voltage
applied to any terminal
1000 V ac rms or dc
Storage temperature -40 °C to 60 °C (-40 °F to 140 °F)
Operating temperature -20 °C to 55 °C (-4 °F to 131 °F)
Temperature coefficient 0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F)
Relative humidity,
non-condensing
< °C
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Vibration Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock 1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor)
Electromagnetic
compatibility
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: accuracy =
specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997)
Safety Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and
IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V
Certifications CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard EN 61010 Part 1-1002
Batteries Four AA batteries (NEDA 15A or IEC LR6)
Battery life Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests
with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into
1 MΩ with a duty cycle of 5 seconds on and 25 seconds off
Size 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight 550 g (1.2 lb)
IP rating IP40
Altitude (operating) 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V
Storage 12,000 m
Over-range capability 110 % of range except for capacitance which is 1 %
Compliance to EN 61557 IEC61557-1, IEC61557-2