User Manual

1587/1577
Users Manual
30
Altitude ................................................................Operating: 2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V,
CAT III 600 V
Storage................................................................ 12,000 m
Over-Range Capability........................................ 110% of range except for capacitance which is 1 %
Compliance to EN 61557 ....................................IEC61557-1, IEC61557-2
AC Voltage Measurement
1587 Accuracy
Range Resolution
50 Hz to 60 Hz
± (% of Rdg + Digits)
60 Hz to 5000 Hz
± (% of Rdg + Digits)
600.0 mV 0.1 mV + (1 % + 3) + (2 % + 3)
6.000 V 0.001 V + (1 % + 3) + (2 % + 3)
60.00 V 0.01 V + (1 % + 3) + (2 % + 3)
600.0 V 0.1 V + (1 % + 3) + (2 % + 3)
1
1000 V 1 V + (2 % + 3) + (2 % + 3)
1
1. 1 kHz bandwidth.
1587 Lowpass Filter Voltage
Range Resolution
50 Hz to 60 Hz
± (% of Rdg + Digits)
60 Hz to 400 Hz
± (% of Rdg + Digits)
600.0 mV 0.1 mV + (1 % + 3) + (2 % + 3)
6.000 V 0.001 V + (1 % + 3) + (2 % + 3)
60.00 V 0.01 V + (1 % + 3) + (2 % + 3)
600.0 V 0.1 V + (1 % + 3) + (2 % + 3)
1000 V 1 V + (2 % + 3) + (2 % + 3)