Input Impedance .................................................10 MΩ (nominal), < 100 pF
Normal Mode Rejection Ratio .............................> 60 dB @ 50 Hz or 60 Hz
Common Mode Rejection Ratio ..........................>120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)
Overload Protection ............................................ 1000 V rms or dc
DC Millivolts Measurement
Range Resolution
Accuracy 1587
± (% of Rdg + Digits)
Accuracy 1577
± (% of Rdg + Digits)
600.0 mV dc 0.1 mV + (0.1 % + 1) + (0.2 % + 1)
DC and AC Current Measurement
Range Resolution
Accuracy 1587
+
(% of Rdg+Digits)
Accuracy 1577
+ (% of Rdg+Digits)
Burden Voltage
(Typical)
400 mA .1 mA + (1.5 % + 2)
1
+ (2 % + 2)
1
AC
45 Hz to 1000 Hz
60 mA .01 mA +
(1.5 % + 2)
1
+ (2 % + 2)
1
2 mV/mA
400 mA .1 mA + (0.2 % + 2) + (1.0 % + 2) DC
60 mA .01 mA +
(0.2 % + 2) + (1.0 % + 2)
2 mV/mA
1. 1 kHz bandwidth.
Overload ............................................................600 mA for 2 minutes maximum
Overload Protection ............................................ 440 mA, 1000 V, FAST fuse
AC Conversion ....................................................Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are
true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor
can be up to 3 at full scale up to 300 mA, decreasing linearly to crest factor <= 1.5 at
600 mA. For non-sinusoidal waveforms add +(2 % reading + 2 % FS) typical, for a crest
factor up to 3.