Data Sheet

Technical Data
Current clamp set
20 A/200 A 1-, 3-phase
and 3-phase+N
Electrical specifications
Nominal current |n 20 / 200 A ac r.m.s.
Overload Up to 260 A ac r.m.s.
Max. non-destructive current 300 A ac r.m.s.
Conductor position influence < 0.5 % of range at 50/60 Hz
Error due to adjacent conductor 15 mA/A at 50 Hz
Phase shift at 2 kHz at 200 A < ± 1 degree
Frequency range 40 Hz to 5 kHz (- 3 dB)
Temperature coefficient 0.015 % of range/°C
Working voltage (see section safety) 600 V ac r.m.s. or dc
Accuracy (at 50 Hz)
Primary current 0.5 A to 10 A 10 A to 50 A 50 A to 100 A 100 A to 200 A
Operating error in % of reading ± 2.5 % ± 1.5 % ± 0.5 % ± 0.5 %
Phase shift Not specified < 2° < 0.5° < 1°
These current clamp sets have been
designed for applications with Fluke
1735 and Fluke 1740 series products for
accurate non intrusive measurement of
ac currents. Using latest technologies
(integrated memory for calibration data)
provides current ranges from 0.5 A up to
200 A in a frequency range of 40 Hz to
5 kHz. The clamps are approved for the
required installation category.