® 1750 Power Recorder Calibration Manual November 2007 Rev. 1, 12/08 © 2007-2008 Fuke Corporation, All rights reserved. Product specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Introduction........................................................................................................ Contacting Fluke................................................................................................ Safety Information ............................................................................................. Symbols .............................................................................................................
1750 Calibration Manual ii
List of Tables Table 1. 2. 3. 4. 5. 6. 7. 8. Title Symbols.................................................................................................................. Required Equipment............................................................................................... Required Cables ..................................................................................................... AC Voltage Accuracy Verification ........................................................................
1750 Calibration Manual iv
List of Figures Figure 1. 2. 3. 4. 5. 6. Title Power View Software Live View........................................................................... Ground Calibration Cable ...................................................................................... Rogowski Calibration Cable .................................................................................. Rogowski Resistor Measurement Cable ................................................................ Voltage Calibration Cable..........
1750 Calibration Manual vi
1750 Power Recorder Introduction XW Warning To avoid electric shock or personal injury, do not perform the calibration verification tests or calibration procedures described in this manual unless you are qualified to do so. The information provided in this manual is for the use of qualified personnel only.
1750 Calibration Manual To contact Technical Support: fpqsupport@fluke.com or 888-257-9897 (US only) Safety Information The 1750 Power Recorder (hereafter referred to as the Recorder) complies with EN 61010-1 600 V CAT IV, 1000 V CAT III overvoltage rating. See Specifications for a full agency compliance list. XWWarning To avoid electrical shock, personal injury, or fire: 2 • Review the entire manual before using the Recorder and its accessories and observe all warnings and cautions.
Power Recorder Symbols Symbols Symbols used in this manual and on the Recorder are listed in Table 1. Table 1. Symbols X Hazardous voltage. Risk of electrical shock. W Risk of danger. Important information. See manual. P Conforms to requirements of European Union and European Free Trade Association (EFTA). ~ Do not dispose of this product as unsorted municipal waste. Go to Fluke’s web site for recycling information. ) Canadian Standards Association. [ Note: Canadian and US. ] .
1750 Calibration Manual Synchronization and Sampling PLL-Synchronization Source........................... The PLL synchronizes to the A-N voltage for wye power types, and to the A-B voltage for delta power types. All listed power types can be characterized as either wye or delta. PLL Lock Range................................................ 42.5 to 69 Hz Sampling Frequency ........................................
Power Recorder Specifications for the System: Recorder and Power Analyze Software Measurement Accuracy ...................................... Same as rms voltage Voltage Dip (RMS sag) Measurement Type ............................................. True rms (one cycle calculation by overlapping each half cycle) (voltage between lines is measured for 3P3W lines and phase voltage is measured for 3P4W lines) Displayed Data ...................................................
1750 Calibration Manual Environmental and Safety Specifications Operating Environment ....................................... indoors or in covered area outdoors, up to 2000 m altitude Storage Temperature and Humidity.................... -20 °C to 50 °C, 80 % rh max, non-condensing Operating Temperature and Humidity ................ 0 °C to 40 °C, 80 % rh max, non-condensing Maximum Rated Working Voltage Voltage Terminals ............................................. 1100 V rms Voltage Durability..........
Power Recorder Required Equipment The cables listed in Table 3 are necessary for verification and calibration adjustment. To assemble the cables, see cable assembly figures in the Test Cables section. Table 3.
1750 Calibration Manual Verification Tests XWWarning To avoid electrical shock, personal injury, or fire: • Do not perform the calibration procedures or calibration verification tests described in this manual unless you are qualified to do so. • Repairs or servicing should be performed only by qualified personnel. The following tests are used to verify the functions of the Recorder. If the Recorder fails any of the verification tests, calibration or repair may be needed.
Power Recorder Verification Tests power analyze live view screen.bmp Figure 1. Power View Software Live View 7. Connect the calibrator NORMAL HI to inputs A, B, C, and J of the Recorder. 8. Connect the NORMAL LO to input N and the protective conductor terminal . of the Recorder. 9. Apply the voltages from Table 4 and list the Recorder readings for channels AN, BN, CN, and NG in the appropriate areas. Sine waves will be displayed for AN, BN and CN, all at the same amplitude and phase shift.
1750 Calibration Manual Neutral Voltage Verification 1. Connect NORMAL LO to A, B, C, J, and .. 2. Connect the calibrator NORMAL HI to N on the Recorder. 3. Apply the voltages from Table 5, listing the Recorder readings for the appropriate model. Table 5. Neutral (AC) Voltage Accuracy Check Applied NG Reading Limits 60 V @ Local Hz ± 1.12 V 100 V @ Local Hz ± 1.20 V 4. Verify that the N LED is on. 5. Set the calibrator to STBY, verify that the LED for N turns off. 6.
Power Recorder Verification Tests Table 6. Rogowski Current Accuracy Check Applied Expected Expected Voltage Current Reading A Reading B Reading C Reading N Reading G Reading Limits 0.649 V 6.49 μA 20 A ± 0.20 A 1.6225 V 6.225 μA 50 A ± 0.35 A 3.245 V 32.45 μA 100 A ± 0.60 A CT Current Verification 1. Connect the calibrator NORMAL HI to Voltage A, B, and C on the Recorder. 2. Connect the calibrator NORMAL LO to Voltage N and J on the Recorder. 3.
1750 Calibration Manual Watt Verification 1. Use the same connections as the CT current verification and apply 440 V and local line frequency out of the NORMAL output and 1.9 V on the AUX output. 2. Click on the power button in Power Analyze. The A, B and C should be 41.8 kW ± 0.3 kW. 3. On the calibrator press the blue button for Phase and Ref Menu. 4. Press the blue button for AUX Phase NRM. 5. Enter 90 with the numerical keypad and press Enter. The power should be 0 ± 0.1 kW.
Power Recorder Verification Tests 17. Set the calibrator to 1000 V 10 kHz. 18. Connect the voltage test cable to UUT voltage input J for approximately 15 seconds, then disconnect. Note, this signal is designated G in Power Analyze. 19. Press STBY on the calibrator. 20. Set the calibrator to 0 V 0 Hz. To Use Power Analyze to Verify the Amplitude Accuracy of Recorded Impulse Events 21. Wait two minutes for the UUT to compile event information. 22. Click on 1750 Download, and then Save. 23.
1750 Calibration Manual Table 8. Impulse Verification Inputs Applie d Voltage [Vrms] Expected Reading [Vpk] Channel Measured Magnitude |Vpk| Limits [Vpk] 800 1131 AN (A) ± 77 850 1202 BN (B) ± 80 900 1273 CN (C) ± 84 950 1344 NG (N) ± 87 1000 1414 NG (G) ± 91 Calibration Required Equipment The required equipment and cables for calibrating the Recorder are listed in Tables 2 and 3.
Power Recorder Calibration Calibration is controlled by 1750 Cal Wizard Software (Cal Wizard), which communicates with the Recorder over Ethernet. Cal Wizard steps through the calibration procedure, with instructions for making cable connections, setting the correct sources, and finalizing the calibration. The following steps detail the calibration adjustment procedure: 1. Connect the Recorder and the PC to the network with an Ethernet cable.
1750 Calibration Manual Test Cables The following figures detail the assembly of cables used for calibrating the Recorder. fdo02.eps Figure 2.
Power Recorder Test Cables fdo01.eps Figure 3.
1750 Calibration Manual fdo05.eps Figure 4.
Power Recorder Test Cables fdo03.eps Figure 5.
1750 Calibration Manual fdo04.eps Figure 6.