75, 177, & 179 True RMS Multimeters Calibration Information Introduction XWWarning To avoid electric shock or injury, do not perform the performance tests or calibration adjustment procedures unless qualified to do so. The information provided in this document is for the use of qualified personnel only.
175, 177, & 179 Calibration Information Safety Information "Warning" and "Caution" Statements A “XW Warning" identifies hazardous conditions and actions that could cause bodily harm or death. A "W Caution" identifies conditions and actions that could damage the Meter, the equipment under test, or cause permanent loss of data.
True RMS Multimeters International Electrical Symbols International Electrical Symbols The following international symbols appear in this document and on the Meter. Earth ground AC or DC J I P W Risk of danger. Important Information. Refer to the manual. $ Conforms to relevant Canadian Standards Association directives X Hazardous Voltage T s Technischer Überwachungs Verein. German certifying body. ! b Battery. Low Battery when shown on display.
175, 177, & 179 Calibration Information Specifications Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications take the form of: ±( [ % of Reading ] + [ Counts ] ). AC conversions are ac-coupled, true rms responding.
True RMS Multimeters Specifications Function AC Volts 2,3 Range 1 Resolution Model 175 Model 177 Model 179 600.0 mV 0.1 mV 1.0 % + 3 1.0 % + 3 1.0 % + 3 6.000 V 0.001 V (45 Hz to 500 Hz) (45 Hz to 500 Hz) (45 Hz to 500 Hz) 60.00 V 0.01 V 600.0 V 0.1 V 1000 V 1V 2.0 % + 3 2.0 % + 3 2.0 % + 3 (500 Hz to 1 kHz) (500 Hz to 1 kHz) (500 Hz to 1 kHz) 0.15 % + 2 0.09 % + 2 0.09 % + 2 0.15 % + 2 0.09 % + 2 0.09 % + 2 0.15 % + 2 0.15 % + 2 0.15 % + 2 DC mV 600.0 mV 0.
175, 177, & 179 Calibration Information Function DC Amps 5 Range 60.00 mA 400.0 mA 6.000 A 10.00 A 1 Model 175 Model 177 Model 179 0.01 mA 6 0.1 mA 0.001 A 7 1.0 % + 3 1.0 % + 3 1.0 % + 3 0.1 % + 1 0.1 % + 1 0.1 % + 1 NA NA 0.01 A Hz 99.99 Hz 0.01 Hz (AC- or DC- 999.9 Hz 0.1 Hz coupled, 9.999 kHz 0.001 kHz V or A 99.99 kHz 0.01 kHz -40 °C to +400 °C 0.1 °C -40 °F to +752 °F 0.
True RMS Multimeters Testing the Fuses Testing the Fuses XWWarning To avoid electrical shock or personal injury: • Remove the test leads and any input signals before replacing the battery or fuses. • To avoid electrical shock, arc blast, or damage to the Meter, install only fuses with the amperage, interrupt, voltage, and speed ratings specified in Table 5. To test the fuses (refer to Figure 1): 1. Set the Rotary Switch to e. 2.
175, 177, & 179 Calibration Information Replacing the Fuses XWWarning To avoid electrical shock, arc blast, or damage to the Meter, install only fuses with the amperage, interrupt, voltage, and speed ratings specified in Table 5. To replace the Fuses (refer to Figure 2): 1. Turn the Rotary Switch to OFF. 2. Disconnect the test leads and/or any connectors from the terminals. 3. Remove the four screws from the Case Bottom and separate the Case Top from the Case Bottom. 4.
True RMS Multimeters Cleaning Cleaning XWWarning To avoid electrical shock: • Remove test leads and any input signals before cleaning. • Do not reinstall the pca until it is completely dry. Dirt or moisture in the terminals can affect readings. W Caution To avoid damaging the Meter, do not use aromatic hydrocarbons or chlorinated solvents for cleaning. These solutions will react with the plastics used in the instruments. Do not use detergents of any kind for cleaning the pca.
175, 177, & 179 Calibration Information LCD Mask aik15f.eps Figure 3. Removing the LCD Mask Performance Tests XWWarning To avoid electric shock, do not perform the performance test procedures unless the Meter is fully assembled. The following performance tests verify the complete operation of the Meter and check the accuracy of each Meter function against its specifications. If the Meter fails any part of the test, calibration adjustment and/or repair is indicated.
True RMS Multimeters Performance Tests Required Equipment A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures in this document. If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1. Table 1. Calibrator Specifications Recommended Equipment 5500A Multi-Product Calibrator (or equivalent) Measurement Function Accuracy DC Volts 30 mV to 1000 V ±0.0225 % DC Current 3 mA to 9 A ±0.
175, 177, & 179 Calibration Information Performance Tests Note When calibrating or measuring AC current, avoid the potential for errors from coupled noise, by either: 1. Shorting the VeG to the COM input or 2. Having NO connection to the VeG input. 1. Set the UUT Rotary Switch to the position called for in Table 2 (175) or Table 3 (177/179), and for all measurements other than current (amps), connect the calibrator to the VeG and COM input terminals.
True RMS Multimeters Performance Tests Table 2. Performance Tests 175 (continued) Meter Response Test (Switch Position) L 5500 Output Lower Limit Upper Limit 5V 4.990 V dc 5.010 V dc 300 V 299.3 V dc 300.7 V dc 1000 V 996 V dc 1004 V dc -1000 V -1004 V dc -996 V dc D Hz 3 V 45 Hz 44.94 Hz 45.06 Hz DC Volts Frequency 1 30 V 50 kHz 49.94 kHz 50.06 kHz E 30 mV 0 Hz 29.8 mV dc 30.2 mV dc DC Millivolts -300 mV -300.7 mV dc -299.3 mV dc 600 mV 598.9 mV dc 601.
175, 177, & 179 Calibration Information Table 3. Performance Tests 177/179 Test K AC Volts 5500 Output Meter Response Lower Limit Upper Limit 300 mV 45 Hz 296.7 mV ac 303.3 mV ac 5 V 500 Hz 4.947 V ac 5.053 V ac 5 V 1 kHz 4.897 V ac 5.103 V ac 50 V 45 Hz 49.47 V ac 50.53 V ac 50 V 1 kHz 48.97 V ac 51.03 V ac 300 V 45 Hz 296.7 V ac 303.3 V ac 500 V 500 Hz 494.7 V ac 505.3 V ac 500 V 1 kHz 489.7 V ac 510.3 V ac 1000 V 45 Hz 987 V ac 1013 V ac 1 V 45 Hz 44.94 Hz 45.
True RMS Multimeters Calibration Adjustment Table 3. Performance Tests 177/179 (continued) Test F 5500 Output Meter Response Lower Limit Upper Limit 3 mA 45 Hz 2.92 mA ac 3.08 mA ac 50 mA 1 kHz 49.22 mA ac 50.78 mA ac 400 mA 1 kHz 393.7 mA ac 406.3 mA ac J 4.0 A 45 Hz 3.937 A ac 4.063 A ac AC Amps 9.0 A 1 kHz 8.83 A ac 9.17 A ac mI 3 mA, 0 Hz 2.94 mA dc 3.06 mA dc 50 mA 49.47 mA dc 50.53 mA dc -400 mA -404.3 mA dc -395.7 mA dc 4.0 A 3.957 A dc 4.043 A dc -9.0 A -9.
175, 177, & 179 Calibration Information Calibration Adjustment Procedure Use the following steps to adjust the Meter’s calibration: For Meters with a serial number of 84060000 and below, switch the Meter to volts AC (C). For Meters with a serial number of 83911001 – 83911240 and 84060001 and above, switch the Meter to E DC. 1. Turn the Meter over and find the Calibration Seal located near the top of the Meter (refer to Figure 4). 2.
True RMS Multimeters Calibration Adjustment Calibration Seal Calibration Button aik14f.eps Figure 4. Calibration Access Table 4. Calibration Steps If the Meter has a serial number of 84060000 or below, or 83911241 to 84060000, use the following steps. Function (Switch Position) C Calibration Step Input Value C-01 6.000 V, 900.0 Hz C-02 600.0 mV, 100 Hz C-03 6.000 V, 100 Hz C-04 6.000 V dc C-05 60.00 V dc C-06 600.0 V dc C-07 1000 V dc C-08 600.0 mV dc C-09 0.
175, 177, & 179 Calibration Information Table 4. Calibration Steps (continued) If the Meter has a serial number from 83911001 to 83911240 or 84060001 and above, or if the model number flashes on the display at turn on, use the following steps.* Function (Switch Position) E Calibration Step Input Value C-01 600.0 mV dc C-02 120.0 mV dc C-03 6.000 V dc C-04 60.00 V dc C-05 600.0 V dc C-06 600.0 mV, 60 Hz (AC Volts) C-07 600.0 V, 60 Hz e (Ohms) C-08 600.0 Ω C-09 6.000 kΩ C-10 60.
True RMS Multimeters User-Replaceable Parts and Accessories User-Replaceable Parts and Accessories User-replaceable parts and accessories are shown in Table 5 and Figure 5. Table 5.
175, 177, & 179 Calibration Information 16 17 2X 15 4X 14 18 12 13 6 5 4 3 24 21 5X 2 22 11 19 20 1 10 23 9 8 7 aik16f.eps Figure 5.
True RMS Multimeters User-Replaceable Parts and Accessories Lifetime Limited Warranty Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As used herein, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty period shall be at least ten years from the date of purchase.
175, 177, & 179 Calibration Information 22