Data Sheet

Page 1 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
©2000 Fluke Corporation
Rev. A-11/2000
Fluke 170 Series
True-rms Digital Multimeter
Extended Specifications
Model Differences
(all other specifications are the same for each model)
Feature Model
175 177 179
Backlight ••
Temperature Measurement
Basic dc voltage accuracy 0.15% 0.09% 0.09%
Function Absolute Range or Description
AC Voltage, True-rms 0.1 mV to 1000V (1 kHz)
DC Voltage 0.1 mV to 1000V
Continuity Beeper guaranteed on < 25 , guaranteed off > 250 ;
detects opens or shorts of 250 µs or longer.
Resistance 0.1 to 50.00 M
Diode Test 2.400V
Capacitance 1 nF to 9999 uF
AC Current, True-rms 0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
DC Current 0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
Frequency 2 Hz to 50 kHz
Temperature (179 Only) -40 ºC to +400 ºC; -40 ºF to +752 ºF
Basic dc voltage accuracy 0.15% (175) 0.09% (177 & 179)
Basic ac voltage accuracy 1.0%
Features
Feature Description
Digital Display 6000 counts, updates 4 x second
Analog Bargraph Display 33 segments, updates 40 x second
Backlight (177 & 179 only) Automatically turns off after 2 minutes to save battery life
The timeout feature can be disabled with a power-up option
HOLD & Auto HOLD HOLD: freezes the display at the push of a button
Auto HOLD: Display holds present reading until it detects new stable input,
then the meter beeps and displays new reading
MIN MAX AVG Minimum, maximum, and average reading memory
Manual or auto ranging In auto range, the meter selects the range with the best resolution for
the present measurement value
Fast continuity/open detection The beeper sounds with a stretched pulse for opens or shorts as
brief as 250 µs
Test lead alert The message “LEAd” appears briefly on the display when the rotary switch
is moved to or from any A (Amps) position
Power-up options (1) Turn on all LCD segments, (2) Disable beeper, (3) Disable sleep mode,
(4) Enable smoothing, (5) Disable backlight timeout (Models 177 & 179 only)
Closed-case calibration No internal adjustments needed
Probe holders The instrument comes with built-in probe holders for probe storage and
for convenience when making measurements
Battery access door Battery replacement without voiding calibration
High-impact overmold case Integrated overmolded protection provides superior impact protection for
your meter
Nominal Specifications

Summary of content (3 pages)