User's Manual

Performance Verification
Scope Input A, B, C, D Tests 4
4-25
Table 4-11. Input B Trigger Sensitivity Test Points
UUT
Model
UUT
Time base
5502A SC... MODE levsin UUT
Trigger Amplitude
Initial Input Voltage Frequency
190-502 200 ns/div 100 mV pp 5 MHz 0.5 div
190-062
10 ns/div 400 mV pp 60 MHz 1 div
10 ns/div 800 mV pp 100 MHz 2 div
190-102
190-104
10 ns/div 400 mV pp 100 MHz 1 div
10 ns/div 800 mV pp 150 MHz 2 div
190-202
190-204
10 ns/div 400 mV pp 200 MHz 1 div
10 ns/div 800 mV pp 250 MHz 2 div
190-502
190-504
2 ns/div 400 mV pp 500 MHz 1 div
2 ns/div 800 mV pp 600 MHz 2 div
Input B AC Voltage Accuracy (HF) and Bandwidth Test
Proceed as follows to test the Input B high frequency automatic scope ac voltage
measurement accuracy and the bandwidth:
1. Connect the Test Tool to the 5502A as shown in Figure 4-8.
2. Select the Test Tool setup:
a) Recall the setup (see Standard Test Tool Setup). Press
SAVE
,
F2
(RECALL) and select SETUP, press
ENTER
, select the setup name, and
press
ENTER
to recall the setup.
b) Press
SCOPE
, then press
F2
(READING…), select
F1
(
READINGS 2), and select on B | V ac.
c) Press
MANUAL
AUTO
to select autoranging (AUTO in upper right LCD edge).
d) Use
mV
RANGE
and
RANGE
to change the sensitivity range to select manual
sensitivity ranging and lock the Input B sensitivity range on 500 mV/div.
e) Press
TRIGGER
and use
F1
to select Input B as trigger source.
3. Set the 5502A to source a sine wave at the first test point in Table 4-12.
4. Observe the Input B reading and check that it is within the range shown
under the appropriate column of Table 4-12.
5. Continue through the test points.