User's Manual

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
4-32
9. Use
mV
RANGE
and
RANGE
to set the Input A sensitivity to 2 V/div (the actual probe
setting is 10:1).
10. Use
TIMEsns
to select the time base to 20 μs/div.
11. Set the calibrator to video mode with amplitude +100 %. Set format and
marker line number to:
PAL 622 (even) for PAL and PALplus
SECAM 622 (even) for SECAM
NTSC 262 even for NTSC
12. Observe the trace and check if the Test Tool triggers on the negative pulse
before the marker pulse (see Figure 4-12).
13. Use
to select Test Tool line number:
310 for PAL, PALplus, or SECAM
262 for NTSC
14. Set the calibrator format and marker line number to:
PAL 310 (odd) for PAL and PALplus
SECAM 310 (odd) for SECAM
NTSC 262 odd for NTSC