Datasheet
2 Fluke Corporation 1587/1577 Insulation Multimeters
AC voltage measurement
1587 accuracy
Range Resolution 50 Hz to 60 Hz 60 Hz to 5000 Hz
600.0 mV 0.1 mV + (1 % + 3) + (2 % + 3)
6.000 V 0.001 V + (1 % + 3) + (2 % + 3)
60.00 V 0.01 V + (1 % + 3) + (2 % + 3)
600.0 V 0.1 V + (1 % + 3) + (2 % + 3)
1
1000 V 1 V + (2 % + 3) + (2 % + 3)
1
1
1 kHz bandwidth
1587 lowpass filter Voltage
Range Resolution 50 Hz to 60 Hz 60 Hz to 400 Hz
600.0 mV 0.1 mV + (1 % + 3) + (2 % + 3)
6.000 V 0.001 V + (1 % + 3) + (2 % + 3)
60.00 V 0.01 V + (1 % + 3) + (2 % + 3)
600.0 V 0.1 V + (1 % + 3) + (2 % + 3)
1000 V 1 V + (2 % + 3) + (2 % + 3)
1577 accuracy
Range Resolution 50 Hz to 60 Hz
600.0 mV 0.1 mV + (2 % + 3)
6.000 V 0.001 V + (2 % + 3)
60.00 V 0.01 V + (2 % + 3)
600.0 V 0.1 V + (2 % + 3)
1000 V 1 V + (2 % + 3)
AC conversion: Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are
true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor can be up to 3 at
full scale up to 500 V, decreasing linearly to crest factor < = 1.5 at 1000 V. For non-sinusoidal waveforms
add ± (2 % reading + 2 % FS) typical, for a crest factor up to 3.
Input impedance: 10 MΩ (nominal), < 100 pF, ac-coupled
Common mode rejection ratio (1 kΩ unbalanced): > 60 dB at dc, 50 or 60 Hz
Overload protection: 1000 V rms or dc, 10
7
V Hz Max
DC voltage measurement
R
ange
R
esolution
Accurac
y 1587
1
Accurac
y 1
5
7
7
1
6.000 V dc 0.001 V 0.09 % + 2 0.2 % + 2
60.00 V dc 0.01 V 0.09 % + 2 0.2 % + 2
6
00.0 V dc 0.1 V 0.09 % + 2 0.2 % + 2
1
000 V dc
1 V
0.0
9 % + 2
0.2 % + 2
1
Accuracies apply to ± 100 % of range
Input impedance: 1
0 M
Ω (nominal), < 1
00 pF
Normal mode rejection ratio: > 60 dB @ 50 Hz or 60 Hz
Common rode rejection ratio: > 120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)
Overload protection: 1
000 V rms or dc
DC millivolts measurement
R
ange
R
esolution
Accurac
y 1
58
7
Accurac
y 1577
600.0 mV dc 0.1 mV 0.1 % + 1 0.2 % + 1
1587/1577 Specifications