Product Manual

291, 292, 294
Users Manual
16-4
The full procedure (for the 4-channel instrument) is as follows:
CAL 03 CH1 DC offset zero. Adjust for 0 V ±5 mV
CAL 04 CH1 DC offset at + full scale. Adjust for +10 V ±10 mV
CAL 05 CH1 DC offset at full scale. Check for –10 V ±3 %
CAL 06 CH1 Multiplier zero. Adjust for minimum volts ac
CAL 07 CH1 Multiplier offset. Adjust for 0 V ±5 mV
CAL 08 CH1 Waveform offset. Adjust for 0 V ±5 mV
CAL 09 CH1 Output level at full scale Adjust for 10 V ±10 mV
CAL 10 CH1 20 dB attenuator Adjust for 1 V ±1 mV
CAL 11 CH1 40 dB attenuator Adjust for 0·1 V ±·1 mV
CAL 12 CH1 10 dB attenuator Adjust for 2·236 V ac ±10 mV
CAL 13 CH1 Not used.
CAL 14 CH1 Not used.
CAL 15 CH1 Not used.
CAL 16 CH1 Level 0.1 MHz Note reading
CAL 17 CH1 Level 33 MHz Check reading
CAL 18 CH1 Level 1 MHz Adjust for same reading
CAL 19 CH1 Level 2 MHz Adjust for same reading
CAL 20 CH1 Level 4 MHz Adjust for same reading
CAL 21 CH1 Level 5 MHz Adjust for same reading
CAL 22 CH1 Level 10 MHz Adjust for same reading
CAL 23 CH1 Level 15 MHz Adjust for same reading
CAL 24 CH1 Level 20 MHz Adjust for same reading
CAL 25 CH1 Level 25 MHz Adjust for same reading
CAL 26 CH1 Level 30 MHz Adjust for same reading
CAL 27 CH1 Level 32·5 MHz Adjust for same reading
CAL 28 CH1 Level 35 MHz Adjust for same reading
CAL 29 CH1 Level 37·5 MHz Adjust for same reading
CAL 30 CH1 Level 40 MHz Adjust for same reading
CAL 31 CH2 DC offset zero. Adjust for 0 V ±5 mV
CAL 32 CH2 DC offset at + full scale. Adjust for +10 V ±10 mV
CAL 33 CH2 DC offset at full scale. Check for –10 V ±3 %
CAL 34 CH2 Multiplier zero. Adjust for minimum volts ac
CAL 35 CH2 Multiplier offset. Adjust for 0 V ±5 mV
CAL 36 CH2 Waveform offset. Adjust for 0 V ±5 mV
CAL 37 CH2 Output level at full scale Adjust for 10 V ±10 mV
CAL 38 CH2 20 dB attenuator Adjust for 1 V ±1 mV
CAL 39 CH2 40 dB attenuator Adjust for 0·1 V ±·1 mV
CAL 40 CH2 10 dB attenuator Adjust for 2·236 V ac ±10 mV
CAL 41 CH2 Sum offset Adjust for 0 V ±5 mV
CAL 42 CH2 SCM level at full-scale Adjust for 5 V ±5 mV
CAL 43 CH2 AM level at full-scale Adjust for 10 V ±10 mV
CAL 44 CH2 Level 0.1 MHz Note reading
CAL 45 CH2 Level 33 MHz Check reading
CAL 46 CH2 Level 1 MHz Adjust for same reading
CAL 47 CH2 Level 2 MHz Adjust for same reading
CAL 48 CH2 Level 4 MHz Adjust for same reading
CAL 49 CH2 Level 5 MHz Adjust for same reading
CAL 50 CH2 Level 10 MHz Adjust for same reading
CAL 51 CH2 Level 15 MHz Adjust for same reading
CAL 52 CH2 Level 20 MHz Adjust for same reading
CAL 53 CH2 Level 25 MHz Adjust for same reading
CAL 54 CH2 Level 30 MHz Adjust for same reading
CAL 55 CH2 Level 32·5 MHz Adjust for same reading